Technical data, Other available ranges – Atec Wallace-Tierman-65-120 User Manual
Page 4
Technical Data
PUBLICATION
TI 650.504DE
Ordering Numbers
The standard scale contains two sets of calibrations (dual scale) in the pressure units inches of water and psi. Dial
markings in the negative range are in red colour and in the positive range in black colour.
Range and Calibration
Ordering No.
Graduation
Outer scale: –100 to 0 to 850
65-120 000
1.0 inch water
inches of water (20 C)
Inner scale: –3.6 to 0 to 30.6 psi
0.05 psi
Other available ranges
Range and Calibration
Ordering No.
Graduation
0 to 4.5
psi (0 to 300 mbar)
65-120 005
0.005 psi (0.5 mbar)
0 to 10
psi (0 to 700 mbar)
65-120 010
0.01
psi (1 mbar)
0 to 15.5 psi (0 to 1
bar)
65-120 015
0.02
psi (0.001 bar)
0 to 20
psi (0 to 1.4 bar)
65-120 020
0.02
psi (0.002 bar)
0 to 30
psi (0 to 2
bar)
65-120 030
0.05
psi (0.002 bar)
0 to 45
psi (0 to 3
bar)
65-120 045
0.05
psi (0.005 bar)
0 to 60
psi (0 to 4
bar)
65-120 060
0.1
psi (0.005 bar)
0 to 100
psi (0 to 7
bar)
65-120 100
0.1
psi (0.01
bar)
Other units of calibration are available at no extra cost. Two sets of calibrations (dual scales) are available at extra
cost.
4
Accuracy
0.066% of full scale
Repeatability
0.03% of full scale
Sensitivity
0.01% of full scale
Hysteresis
0.1% of full scale
Readability
0.02% of full scale
Temperature effect
Maximum is 0.1% of full scale per 10 C change
from 23 C
Scale length
45 inches (1140 mm) through 2 pointer revolutions
Dial diameter
8
1
/
2
inches (215 mm)
Capsule system volume
6.9 cc with pointer at zero; 8.6 cc, at full scale
Case volume
3070 cc
Case pressure
35 psi (2,5 bar) maximum
Overpressure protection
A pressure relief valve protects the mechanism
(capsule); pressure up to 90 psi will not damage the
mechanism nor affect accuracy. A built-in relief valve
and a flow restrictor protect the case from pressures
up to 10 times the case rating of 35 psig. This valve
is an emergency-protective device only.
Wallace & Tiernan GmbH
Postfach 1563, D-89305 Günzburg, GERMANY
Street address: Auf der Weide 10,
D-89312 Günzburg, GERMANY
Tel.: (49) 8221-9040, Fax: (49) 8221-904203
a U.S. Filter Company
Due to continual product development and improvement, certain specifications
may change without prior announcement.
1
2/98
Maximum case leak rate
Will not exceed 1.03 x 10
–7
std cc/sec or 0.018 psi/hr
Materials exposed to measured gas
Capsule system: Ni-Span C; soft solder; brass;
303 stainless steel; silver solder.
Case system: Ni-Span C; brass; phosphor bronze,
beryllium copper, magnesium, aluminum, nylon,
303 stainless steel; Elgiloy; soft solder, silver solder,
Hypalon.
Inlet manifold to pressure gauge; aluminum; brass;
polyethylene; copper; steel; glass wool; carbon;
Buna N.
Gauge case
Anodized aluminum with tempered-glass dial cover
Carrying case
ABS plastic 17
1
/
2
in. x 12 in. x 7
1
/
2
in.
(445 x 305 x 190 mm)
Accessories
6 adapters for
1
/
8
” pipe thread to
1
/
4
” plastic tubing,
20 ft of
1
/
4
” OD plastic tubing, instruction book,
sheet with summary of important instructions
Shipping weight
approx. 26 lb (12 kg); approx. 20 lb (9 kg) net
Note: Calibrator for use with non-corrosive gases.
It should not be used with liquids