beautypg.com

Atec Tektronix-1730 Series User Manual

Page 11

background image

1730–Series Introduction

1–11

Table 1–7: CRT Display

Characteristic

Requirement

Supplemental Information

CRT viewing area

80 X 100 mm
Horizontal = 12.5 div
Vertical = 170 IRE units (1.19 V)

Accelerating potential

Nominally 13.75 kV

Trace rotation range

Greater than

±

1

°

from horizontal

Total adjustment range is typically 8

°

Graticule

Internal, variable illumination

Table 1–8: Power Source

Characteristic

Requirement

Supplemental Information

Mains voltage range

90 – 250 V

Continuous range from 90 to 250 VAC

Mains frequency

50 or 60 Hz

Power consumption

56 VA (35 Watts)

Table 1–9: Environmental Characteristics

Characteristic

Requirement

Operating temperature

0 to 50

°

C (+32 to 122

°

F)

Storage temperature

–40 to 75

°

C (–40 to 158

°

F)

Operating altitude

To 15,000 feet (4572 meters)

Storage altitude

To 50,000 feet (15,240 meters)

Vibration

5 minutes at 5 - 15 Hz with 0.060 inch displacement. 5 minutes at
15 - 25 Hz with 0.040 inch displacement. 5 minutes at 25 - 55 Hz
with 0.020 inch displacement. Military Specification:
Mil-T-28800D, Paragraph 1.2.2, Class 3.

Mechanical shock

Non operating: 50 g’s 1/2 sine, 11 ms duration, 3 shocks per
surface (18 total).

Transportation

Qualified under NSTA Test Procedure 1A, Category II (24 inch
drop).

Humidity

Will operate at 95% relative humidity for up to five days. Do not
operate with visible moisture on the circuit boards.