beautypg.com

Frequency synthesizers, signal generators – Atec Wiltron-Anritsu-69000A-68000B Series User Manual

Page 6

background image

FREQUENCY SYNTHESIZERS, SIGNAL GENERATORS

232

Waveforms

Sinusoid, squarewave, triangle, positive ramp, negative ramp, Gaussian noise, uniform noise,
user defined (Option 10)

Internal AM

Rate

0.1 Hz to 1 MHz sinusoidal, 0.1 Hz to 100 kHz squarewave, triangle, ramps

generator

Resolution

0.1 Hz

Accuracy

Same as instrument timebase

Output

BNC connector, rear panel

External FM input

Front or rear panel BNC, 50

or 600

input impedance

All options selectable from modulation menu

Continuously variable from ±10 kHz per volt to ±20 MHz per volt (locked, locked low noise and unlocked

FM sensitivity

narrow modes), or ±100 kHz per volt to ±100 MHz per volt (unlocked wide mode)
For 500 MHz units, maximum sensitivity is divided by 2 from 1 to 2.2 GHz and is divided by 4 from 500
MHz to 1 GHz.

Unlocked wide: ±100 MHz, DC to 100 Hz rates

Deviation

Unlocked narrow: ±10 MHz, DC to 8 MHz rates

Frequency

Locked: The lesser of ±10 MHz or rate x 300, 1 kHz to 8 MHz rates

modulation

Locked low noise: The lesser of ±10 MHz or rate x 3, 50 kHz to 8 MHz rates

Unlocked wide: DC to 100 Hz

FM bandwidth (3 dB)

Unlocked narrow: DC to 10 MHz
Locked: 1 kHz to 10 MHz
Locked low noise: 30 kHz to 10 MHz

Flatness

±1 dB (10 kHz to 1 MHz rates)

Accuracy

10% (5% typical, ±200 kHz deviation, 100 kHz rate)

Incidental AM

<2% (±1 MHz deviation, 1 MHz rate)

Harmonic distortion

<1% (±1 MHz deviation, 10 kHz rate)

Maximum input

±1 V

Waveforms

Sinusoid, squarewave, triangle, positive ramp, negative ramp, Gaussian noise, uniform noise, user defined
(Option 10)

Internal FM

Rate

0.1 Hz to 1 MHz sinusoidal, 0.1 Hz to 100 kHz squarewave, triangle, ramps

generator

Resolution

0.1 Hz

Accuracy

Same as instrument timebase

Output

BNC connector, rear panel

Narrow mode (DC to 8 MHz rates): The lesser of ±3 radians or ±5 MHz/rate

øM deviation

Wide mode (DC to 1 MHz rates):

The lesser of ±400 radians or ±10 MHz/rate. For 6XXX5 units, maximum deviation is divided by 2 from
>1.0 to

2.2 GHz and is divided by 4 from

0.5 to

1.0 GHz.

øM bandwidth (3 dB,

Narrow mode: DC to 10 MHz

relative to 100 kHz rate)

Wide mode: DC to 1 MHz

øM flatness (relative to

Narrow mode (DC to 1 MHz rates): ±1 dB

Phase

100 kHz rate)

Wide mode (DC to 500 kHz rates): ±1 dB

modulation

øM accuracy

10% (at 100 kHz sine wave)

(øM, Option 6)

External øM input

Front or rear panel BNC (shares the FM input), 50

or 600

input impedance. All options selectable from

modulation menu. Shares connectors with FM.

Continuously variable from ±0.0025 to ±5 radians per volt (narrow øM mode) or ±0.25 to ±500 radians per

External øM sensitivity

volt (wide øM mode), selectable from modulation menu. For 6XXX5 units, maximum sensitivity is divided
by 2 from >1 to

2.2 GHz and is divided by 4 from

0.5 to <1 GHz.

External øM maximum
input

±1 V

Waveforms

Sine, square, triangle, positive ramp, negative ramp, Gaussian noise, uniform noise, user defined (option 10)

Rate

0.1 Hz to 1 MHz for sine wave, 0.1 Hz to 100 kHz for other waveforms

Resolution

0.1 Hz

Accuracy

Same as instrument timebase

Output

BNC connector, rear panel

On/off ratio

>80 dB

Rise/fall time (10 to 90%)

<10 ns (<5 ns typical). (for 6XXX5 units, rise/fall time below 1 GHz is 15 ns)

Minimum levelled pulse

Pulse

width

<100 ns (

2 GHz), <1 µs (<2 GHz)

modulation*

6

Minimum unleveled
pulse width

<10 ns

Pulse overshoot

<10% (for 60 and 65 GHz units, overshoot from 40 to 60 GHz is 20% typical)

Level accuracy relative
to CW

±0.5 dB (

1 µs pulse width), ±1.0 dB (<1 µs pulse width) 100 Hz to 1 MHz PRF

Internal øM
generator
(shares the
internal FM
generator)

69200A/69300A/68200B/68300B modulation

Continued on next page