Preliminary datasheet – Atec Tektronix-PA1000 User Manual
Page 5
Measurement accuracy -
frequency
Frequency (10 Hz to 20 kHz)
0.1% of Reading, with the peak of the signal extending 10% above and 10% below the DC level
Frequency (20 kHz to 1 MHz)
0.1% of Reading, with the peak of the signal extending 25% above and 25% below the DC level
Measurement accuracy - power
Watts accuracy
1
± 0.075% of Reading ± 0.075% of Range (PF=1)
VA accuracy
1
± 0.075% of Reading ± 0.075% of Range
VAR accuracy (typical)
VA ± VA
error
2
- W ± W
error
2
- VA
2
- W
2
PF Accuracy
Cos θ -cos [ θ± ( Vh1
ph.err
± Ah1
ph.err
)] ± 0.002
Measurement accuracy - harmonic
magnitude and phase (typical)
1
Voltage harmonics magnitude
(10 Hz to 1 MHz)
± 0.02% of Reading ± 0.1% of Range ± (0.04*F)% of Reading ± 0.05 V
Voltage harmonics phase
± 0.1 ± [ 0.01 * V
range
/ V
reading
)] ± (0.2 / V
range
) ± ( 0.005 *F )
Current harmonics magnitude
(10 Hz to 1 MHz)
± 0.2% of Reading ± 0.1% of Range ± (0.04*F)% of Reading ± (50 µV /Z
ext
)
Current harmonics phase
± 0.1 ± [0.01 * A
range
/ A
reading
)] ± (0.002 / A
range *
Z
ext
)) ± (0.005 *F)
Physical characteristics
Dimensions
mm
in
Height
102
4.0
Width
223
8.7
Depth
285
11.2
Weight
Kg
lb
Net (without lead set)
3.2
7.0
Temperature
C
F
Operating
0 °C to +40 °C
+32 °F to +102 °F
Nonoperating
-20 °C to +60 °C
-4 °F to +140 °F
1
F is the frequency measured in kHz. In the case of harmonics, F is the harmonic frequency.
Z
ext
is the shunt impedance and must be less than or equal to 10 ohms.
Specifications are valid only when applicable voltage and current inputs are >10% of range. The exception is harmonics where the specification is valid when the magnitude of the harmonic is >2% of range.
Measurement conditions during calibration: Instrument default settings unless otherwise stated, Sine waves applied to V and I inputs, 30 minute warm-up, Temperature 23 °C ±5 °C.
Preliminary Datasheet
www.tektronix.com 5