Specifications, Pa1000 power analyzer – Atec Tektronix-PA1000 User Manual
Page 4
Specifications
All specifications apply to all models unless noted otherwise.
Available measurements
V
rms
- Volts RMS
VTHD - Volts Total Harmonic Distortion
A
rms
- Amps RMS
ATHD - Amps Total Harmonic Distortion
WATT - Watts
Z - Impedance
VA - Volt-Amps
R - Resistance
VAR - Volt-Amps reactive
X - Reactance
FREQ - Frequency
HR - Integrator time
PF - Power factor
WHR - Watt Hours
VPK+ - Volts peak (positive)
VAHrs - VA Hours
VPK- - Volts peak (negative)
VArHr - VAr Hours
APK+ - Amps peak (positive)
AHR - Amp Hours
APK- - Amps peak (negative)
V-harm - Voltage harmonics
VDC - Volts DC
A-harm - Ampere harmonics
ADC - Amps DC
V range
VCF - Voltage crest factor
A range
ACF - Current crest factor
Voltage and current ranges
Voltage ranges
1000 V
peak
, 500 V
peak
, 200 V
peak
, 100 V
peak
, 50 V
peak
, 20 V
peak
, 10 V
peak
Current ranges (20 A shunt)
100 A
peak
, 50 A
peak
, 20 A
peak
, 10 A
peak
, 5 A
peak
, 2 A
peak
, 1 A
peak
, 0.5 A
peak
, 0.2 A
peak
, 0.1 A
peak
Current ranges (1 A shunt)
2.0 A
peak
, 1.0 A
peak
, 0.4 A
peak
, 0.2 A
peak
, 0.1 A
peak
, 0.04 A
peak
, 0.02 A
peak
, 0.01 A
peak
, 0.004 A
peak
, 0.002 A
peak
Measurement accuracy - voltage
Voltage accuracy, V
RMS
(45 Hz to 850 Hz)
1
± 0.05% of Reading ± 0.05% of Range ± 0.05 V
Voltage accuracy, V
RMS
(10 Hz to 45 Hz,
850 Hz to 1 MHz, typical)
± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± 0.05 V (typical)
Voltage accuracy, DC (typical) ± 0.1% of Reading ± 0.1% of Range ± 0.05 V
Effect of common mode
(typical)
100 V, 100 kHz < 500 mV
Measurement accuracy - current
Current accuracy, A
RMS
(45 Hz to 850 Hz)
1
± 0.05% of Reading ± 0.05% of Range ± (50 µV/Z
ext
)
Current accuracy, A
RMS
(10 Hz to 45 Hz,
850 Hz to 1 MHz, typical)
± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± (50 µV/Z
ext
) (typical)
Current accuracy, DC (typical) ± 0.1% of Reading ± 0.1% of Range ± (100 µV/Z
ext
)
Current - peak inrush
accuracy (100 A
peak
range)
2% of Range ± 20 mA
Effect of common mode
(typical)
100 V, 100 kHz, 20 A shunt < 15 mA
100 V, 100 kHz, 1 A shunt < 500 μA
100 V, 100 kHz, external shunt < 40 mV
PA1000 Power Analyzer
4 www.tektronix.com