beautypg.com

Atec Tektronix-TDS-5052-5054-5104 User Manual

Page 6

background image

Waveform Processing/Math

Standard Math

Arithmetic – Add, subtract, multiply and
divide waveforms.

FFT – Magnitude.

Vertical Units – Magnitude: Linear, dB, dBm.

Window Functions – Rectangular, Hamming,
Hanning, Kaiser-Bessel, Blackman-Harris, Gaussian,
Flattop2, Tek Exponential.

Optional Math

Algebraic Expressions – Define extensive algebraic
expressions including waveforms, scalars and results of
parametric measurements e.g. (Integral (Ch1-
Mean(Ch1))*1.414).

Calculus – Integrate, differentiate.

Frequency Domain Functions – Spectral magnitude
and phase, real and imaginary spectra.

Additional Vertical Units – Phase: degrees, radians.

Display Characteristics

Display Type – Liquid crystal active-matrix
color display.

Display Size – 211.2 mm (W) x 158.4 mm (H),
264 mm (10.4 in) diagonal.

Display Resolution – 640 horizontal x 480 vertical
pixels.

Waveform Styles – Vectors, Dots, Intensified
Samples, Variable Persistence, Infinite Persistence.

Computer System
and Peripherals

CPU – Intel Celeron Processor, 800 MHz.

PC System Memory – 128 MB.

Hard Disk Drive – >10 GB capacity.

Floppy Disk Drive – Front panel 3.5 in floppy disk
drive, 1.44 MB capacity.

CD-ROM Drive – Side panel CD-ROM drive.

CD-RW Drive (Optional) – Side panel CD-RW drive.

Printer (Optional) – Built-in thermal printer.

Mouse – Logitech thumb wheel model included, USB
interface.

Keyboard – Order 119-6297-00 (USB interface).

Analysis and Connectivity
Software

TekVISA – Application Programmers Interface (API)
for Windows developers. Documentation includes
descriptions and samples of programming test and
measurement applications on the unit in Visual BASIC,
C and C++.

Plug-and-play Drivers – Provides support to run
National Instrument’s LabVIEW and LabWindows
on an external PC connected to a TDS5000
Series oscilloscope.

TekVISA Control (TVC) – ActiveX controls to make
access to TekVISA easy for integration into Microsoft
Windows applications.

Excel Toolbar – Provides direct access to waveforms
and measurements on the oscilloscope from a toolbar
in Excel.

VXI-11 server – An Application Programmers
Interface (API) for LAN connectivity from non-Windows
environments.

Note: For information regarding TDS5000 Series
compatibility with National Instruments hardware
and software products, contact your local Tektronix
account manager.

Note: For information about using a TDS5000
Series oscilloscope as a GPIB controller, contact
your local Tektronix account manager.

Input/Output Ports

Auxiliary Input – Front panel BNC connector. Trigger
level range is adjustable from +8 V to –8 V. The maxi-
mum input voltage is ±20 V.

Probe Compensator Output – Front panel pins.
Amplitude 1 V ± 1% into a

10 k

load, frequency

1 kHz ±5%.

Analog Signal Output Amplitude – Rear-panel BNC
connector, provides a buffered version of the signal
that is attached to the Channel 3 input. 20 mV/div
±20% into a 1 M

load, 10 mV/div ±20% into a

50

load (4 channel models only).

Analog Signal Output Bandwidth, Typical –
100 MHz into a 50

load (4 channel models only).

Auxiliary Output Levels – Rear-panel BNC connec-
tor, provides a TTL-compatible, negative polarity pulse
when the oscilloscope triggers.

External Reference In – Rear-panel BNC connector.
9.8 MHz to 10.2 MHz.

Parallel Port – IEEE 1284, DB-25 connector.

Audio Ports – Miniature phone jacks for stereo
microphone input and stereo line output.

Trigger System

TDS5052

TDS5054

TDS5104

Sensitivity

Internal DC Coupled

0.35 div DC to 50 MHz

0.35 div DC to

increasing to 1 div at 500 MHz

50 MHz increasing to

1 div at 1 GHz

External (auxiliary input)

400 mV from DC to 50 MHz increasing to 750 mV at 100 MHz

Main Trigger Modes

Auto, Normal and Single

Trigger Sequences

Main, Delayed by time, Delayed by events. All sequences can

include separate horizontal delay after the trigger event to

position the acquisition window in time

Trigger Level Range

Internal

±10 divisions from center of screen

External (auxiliary in)

±8 V

Line

Fixed at 0 V

Trigger Coupling

DC, AC (attenuate <60 Hz), HF reject (attenuate >30 kHz),

LF reject (attenuates <80 kHz), noise reject (reduce sensitivity)

Trigger Holdoff Range

250 ns minimum to 12 s maximum

Digital Phosphor Oscilloscopes

TDS5000 Series

Digital Phosphor Oscilloscopes • TDS5000 Series • www.tektronix.com/tds5000

6

Characteristics, cont.