beautypg.com

Data sheet – Atec Tektronix-MSO4000 Series User Manual

Page 14

background image

Data Sheet

Software

Software

Description

NI LabVIEW

SignalExpress Tektronix

Edition LE

A fully interactive measurement software environment

optimized for the MSO/DPO4000 Series, enables

you to instantly acquire, generate, analyze, compare,

import, and save measurement data and signals using

an intuitive drag-and-drop user interface that does not

require any programming.

Standard MSO/DPO4000 Series support for acquiring,

controlling, viewing, and exporting your live signal data

is permanently available through the software. The full

version (SIGEXPTE) adds additional signal processing,

advanced analysis, mixed signal, sweeping, limit

testing, and user-defined step capabilities and is

available for a 30-day trial period standard with each

instrument.

OpenChoice

®

Desktop

Enables fast and easy communication between a

Windows PC and the MSO/DPO4000 Series, using

USB or LAN. Transfer and save settings, waveforms,

measurements, and screen images. Included Word

and Excel toolbars automate the transfer of acquisition

data and screen images from the oscilloscope into

Word and Excel for quick reporting or further analysis.

IVI Driver

Provides a standard instrument programming

interface for common applications such as LabVIEW,

LabWindows/CVI, Microsoft .NET and MATLAB.

eScope

Enables control of the MSO/DPO4000 Series over a

network connection through a standard web browser.

Simply enter the IP address or network name of the

oscilloscope and a web page will be served to the

browser.

Display Characteristics

Characteristic

Description

Display Type

10.4 in. (264 mm) liquid-crystal TFT color display

Display Resolution

1,024 horizontal × 768 vertical pixels (XGA)

Waveform Styles

Vectors, Dots, Variable Persistence, Infinite

Persistence

Graticules

Full, Grid, Cross Hair, Frame, IRE, and mV

Format

YT and simultaneous XY/YT

Waveform Capture Rate

>50,000 wfm/s

Input/Output Ports

Port

Description

CompactFlash Drive

Front-panel access (Type 1)

USB 2.0 Full-speed Host

Port

Supports USB mass storage devices, printers and

keyboard. Two ports available on rear panel and one

on front panel

USB 2.0 High-speed

Device Port

Rear-panel connector allows for communication/control

of oscilloscope through USBTMC or GPIB

with a TEK-USB-488, and direct printing to all

PictBridge-compatible printers

LAN Port

RJ-45 connector, supports 10/100Base-T

XGA Video Port

DB-15 female connector, connect to show the

oscilloscope display on an external monitor or projector

Auxiliary Input

Front-panel BNC connector. Input Impedance 1 MΩ.

Max input 250 V

RMS

with peaks ±400 V

Probe Compensator

Output

Front-panel pins

Amplitude: 2.5 V

Frequency: 1 kHz

Trigger Out

Rear-panel BNC connector, provides a positive polarity

pulse when the oscilloscope triggers

Kensington Lock

Rear-panel security slot connects to standard

Kensington lock

Power Source

Characteristic

Description

Power Source Voltage

100 to 240 V ±10%

Power Source Frequency 47 to 66 Hz (90 to 264 V)

360 to 440 Hz (100 to 132 V)

Power Consumption

250 W maximum

Physical Characteristics

Dimensions

mm

in.

Height

229

9.0

Width

439

17.3

Depth

137

5.4

Weight

kg

lb.

Net

5

11

Shipping

9.5

22

Rackmount Configuration

5U

Cooling Clearance

2 in. (51 mm) required on left side and rear

of instrument

Environmental

Characteristic

Description

Temperature

Operating

0 ºC to +50 ºC

Nonoperating

–20 ºC to +60 ºC

Humidity

Operating

High: 40 ºC to 50 ºC, 10% to 60% Relative Humidity

Low: 0 ºC to 40 ºC, 10% to 90% Relative Humidity

Nonoperating

High: 40 ºC to 60 ºC, 5% to 60% Relative Humidity

Low: 0 ºC to 40 ºC, 5% to 90% Relative Humidity

Altitude

Operating

3,000 meters (9,843 feet)

Nonoperating

12,000 meters (39,370 feet)

Random Vibration

Operating

0.31 G

RMS

from 5 to 500 Hz, 10 minutes each axis,

3 axes, 30 minutes total

Nonoperating

2.46 G

RMS

from 5 to 500 Hz, 10 minutes each axis,

3 axes, 30 minutes total

Regulatory

Electromagnetic

Compatibility

89/336/EEC

Safety

UL61010-1, Second Edition; CSA61010-1 Second

Edition, EN61010-1:2001; IEC 61010-1:2001

14

www.tektronix.com