Data sheet – Atec Tektronix-MSO4000 Series User Manual
Page 14
Data Sheet
Software
Software
Description
NI LabVIEW
SignalExpress Tektronix
Edition LE
A fully interactive measurement software environment
optimized for the MSO/DPO4000 Series, enables
you to instantly acquire, generate, analyze, compare,
import, and save measurement data and signals using
an intuitive drag-and-drop user interface that does not
require any programming.
Standard MSO/DPO4000 Series support for acquiring,
controlling, viewing, and exporting your live signal data
is permanently available through the software. The full
version (SIGEXPTE) adds additional signal processing,
advanced analysis, mixed signal, sweeping, limit
testing, and user-defined step capabilities and is
available for a 30-day trial period standard with each
instrument.
OpenChoice
®
Desktop
Enables fast and easy communication between a
Windows PC and the MSO/DPO4000 Series, using
USB or LAN. Transfer and save settings, waveforms,
measurements, and screen images. Included Word
and Excel toolbars automate the transfer of acquisition
data and screen images from the oscilloscope into
Word and Excel for quick reporting or further analysis.
IVI Driver
Provides a standard instrument programming
interface for common applications such as LabVIEW,
LabWindows/CVI, Microsoft .NET and MATLAB.
eScope
Enables control of the MSO/DPO4000 Series over a
network connection through a standard web browser.
Simply enter the IP address or network name of the
oscilloscope and a web page will be served to the
browser.
Display Characteristics
Characteristic
Description
Display Type
10.4 in. (264 mm) liquid-crystal TFT color display
Display Resolution
1,024 horizontal × 768 vertical pixels (XGA)
Waveform Styles
Vectors, Dots, Variable Persistence, Infinite
Persistence
Graticules
Full, Grid, Cross Hair, Frame, IRE, and mV
Format
YT and simultaneous XY/YT
Waveform Capture Rate
>50,000 wfm/s
Input/Output Ports
Port
Description
CompactFlash Drive
Front-panel access (Type 1)
USB 2.0 Full-speed Host
Port
Supports USB mass storage devices, printers and
keyboard. Two ports available on rear panel and one
on front panel
USB 2.0 High-speed
Device Port
Rear-panel connector allows for communication/control
of oscilloscope through USBTMC or GPIB
with a TEK-USB-488, and direct printing to all
PictBridge-compatible printers
LAN Port
RJ-45 connector, supports 10/100Base-T
XGA Video Port
DB-15 female connector, connect to show the
oscilloscope display on an external monitor or projector
Auxiliary Input
Front-panel BNC connector. Input Impedance 1 MΩ.
Max input 250 V
RMS
with peaks ±400 V
Probe Compensator
Output
Front-panel pins
Amplitude: 2.5 V
Frequency: 1 kHz
Trigger Out
Rear-panel BNC connector, provides a positive polarity
pulse when the oscilloscope triggers
Kensington Lock
Rear-panel security slot connects to standard
Kensington lock
Power Source
Characteristic
Description
Power Source Voltage
100 to 240 V ±10%
Power Source Frequency 47 to 66 Hz (90 to 264 V)
360 to 440 Hz (100 to 132 V)
Power Consumption
250 W maximum
Physical Characteristics
Dimensions
mm
in.
Height
229
9.0
Width
439
17.3
Depth
137
5.4
Weight
kg
lb.
Net
5
11
Shipping
9.5
22
Rackmount Configuration
5U
Cooling Clearance
2 in. (51 mm) required on left side and rear
of instrument
Environmental
Characteristic
Description
Temperature
Operating
0 ºC to +50 ºC
Nonoperating
–20 ºC to +60 ºC
Humidity
Operating
High: 40 ºC to 50 ºC, 10% to 60% Relative Humidity
Low: 0 ºC to 40 ºC, 10% to 90% Relative Humidity
Nonoperating
High: 40 ºC to 60 ºC, 5% to 60% Relative Humidity
Low: 0 ºC to 40 ºC, 5% to 90% Relative Humidity
Altitude
Operating
3,000 meters (9,843 feet)
Nonoperating
12,000 meters (39,370 feet)
Random Vibration
Operating
0.31 G
RMS
from 5 to 500 Hz, 10 minutes each axis,
3 axes, 30 minutes total
Nonoperating
2.46 G
RMS
from 5 to 500 Hz, 10 minutes each axis,
3 axes, 30 minutes total
Regulatory
Electromagnetic
Compatibility
89/336/EEC
Safety
UL61010-1, Second Edition; CSA61010-1 Second
Edition, EN61010-1:2001; IEC 61010-1:2001
14