Multi Channel Power
Multi Channel Power for an overview of who is on air
MCP mode is perfect for a very fast overview of specified frequency
bands or channels. Service tables can be defined containing up to
500 freely selectable channels each with a dedicated RBW and
service name.
Simultaneous representation of maximum (Max), average (Avg) and
minimum (Min) values allows immediate distinction between
permanent and non-permanent signals.
Level Meter
Level Meter for gapless signal measurements
Level Meter mode allows selective measurements at a defined
frequency (Fcent) e.g. for monitoring a specific channel (Zero Span
operation). Resolution bandwidth (RBW) can be set according to the
channel width in the range of 100 Hz to 32 MHz. The steep filter
characteristics provide precise separation from adjacent channels.
Peak detector values (for short pulsed) and RMS detector values
(for fluctuating signals) are displayed simultaneously. Level Meter
mode provides gapless and interruption-free measurements.
Scope view for detailed analysis versus time
Scope and IQ Demodulation (Option)
Scope mode provides an oscilloscopic view of signals. Almost all
details can be made visible in this mode to allow rapid classification.
The minimum resolution time of 32 ns even allows analysis of high-
speed data transmissions or pulsed signals such as radar.
A top sweep time of 24 hours allows full-day power monitoring of a
single carrier. Extensive trigger functions allow for triggering and
subsequent display of burst signals including a pre-trigger view.
RF signals can be completely described by I/Q data. The I/Q
Demodulation Recorder enables the user to store I/Q data on the
instrument and transfer it to PC for analysis.