beautypg.com

Atec HP-Agilent-8110A User Manual

Page 6

background image

6

Agilent 81107A Multichannel Deskew
for the Agilent 8110A

Can be retrofitted without
recalibration. Supports two
output channels. The
multichannel deskew can be used for
two applications:

Up to 10 channels:

compensates delay

between external input and trigger
outputs when using up to five
8110As synchronously.

Delay calibration:

compensates for

measurement system delays or pre-
trigger delays of oscilloscopes.

Variable range:

0 ns to 28 ns

Additional fixed delay:

6.5 ns typical

Resolution:

10 ps

User Interface
Overprogramming:

all parameters can

be overprogrammed
(exceeding specifications) to fully
exploit the hardware limits.

Setting check:

warning messages indi-

cate potentially conflicting parame-
ters due to inaccuracy.
Error messages indicate
conflicting parameters.

Help key:

displays a context-

sensitive message.

Autoset key:

resolves all timing con-

flicts.

Non-volatile memory:

current

setting is saved on power-down.
Up to nine user settings and one fixed
default setting can be stored in the
instrument.

Clear memory:

clears all nine user set-

tings.

Memory card:

320 settings can be

stored on a 1 MB PCMCIA card
(MS-DOS

).

Remote Control

Operates according to IEEE
standard 488.2, 1987 and SCPI
1992.0.

Function Code:

SH1, AH1, T6, L4, SR1,

RL1, PP0, DC1, DT1, C0.

Programming times:

all checks and

display off.

General

Operating temperature:

0°C to +55°C.

Storage temperature:

-40°C to +70°C.

Humidity:

95% r.h. up to 40°C ambi-

ent temperature.

EMC

:

conforms to EN50082-1,

EN 55011, Class A.

Noise emission:

5.7 bel typical.

Battery:

Lithium CR2477-N.

Safety:

IEC1010, CSA1010.

Power requirements:

100-240 Vac, ± 10%, 50-60 Hz;
100-120 Vac, ± 10%, 400 Hz.

Power consumption:

300 VA max.

Max. dimensions (H * W * D):

89 mm * 426 mm * 521 mm.

Weight:

9.2 kg net, 13.8 kg

shipping.

Recalibration period:

one year recommended.

Warranty:

three years standard.

ASCII command

Typ. exec. time

One parameter

or mode

30 ms typ.

Recall setting

250 ms typ.

16 k pattern transfer

600 ms typ.