Atec HP-Agilent-8110A User Manual
Page 6

6
Agilent 81107A Multichannel Deskew
for the Agilent 8110A
Can be retrofitted without
recalibration. Supports two
output channels. The
multichannel deskew can be used for
two applications:
Up to 10 channels:
compensates delay
between external input and trigger
outputs when using up to five
8110As synchronously.
Delay calibration:
compensates for
measurement system delays or pre-
trigger delays of oscilloscopes.
Variable range:
0 ns to 28 ns
Additional fixed delay:
6.5 ns typical
Resolution:
10 ps
User Interface
Overprogramming:
all parameters can
be overprogrammed
(exceeding specifications) to fully
exploit the hardware limits.
Setting check:
warning messages indi-
cate potentially conflicting parame-
ters due to inaccuracy.
Error messages indicate
conflicting parameters.
Help key:
displays a context-
sensitive message.
Autoset key:
resolves all timing con-
flicts.
Non-volatile memory:
current
setting is saved on power-down.
Up to nine user settings and one fixed
default setting can be stored in the
instrument.
Clear memory:
clears all nine user set-
tings.
Memory card:
320 settings can be
stored on a 1 MB PCMCIA card
(MS-DOS
).
Remote Control
Operates according to IEEE
standard 488.2, 1987 and SCPI
1992.0.
Function Code:
SH1, AH1, T6, L4, SR1,
RL1, PP0, DC1, DT1, C0.
Programming times:
all checks and
display off.
General
Operating temperature:
0°C to +55°C.
Storage temperature:
-40°C to +70°C.
Humidity:
95% r.h. up to 40°C ambi-
ent temperature.
EMC
:
conforms to EN50082-1,
EN 55011, Class A.
Noise emission:
5.7 bel typical.
Battery:
Lithium CR2477-N.
Safety:
IEC1010, CSA1010.
Power requirements:
100-240 Vac, ± 10%, 50-60 Hz;
100-120 Vac, ± 10%, 400 Hz.
Power consumption:
300 VA max.
Max. dimensions (H * W * D):
89 mm * 426 mm * 521 mm.
Weight:
9.2 kg net, 13.8 kg
shipping.
Recalibration period:
one year recommended.
Warranty:
three years standard.
ASCII command
Typ. exec. time
One parameter
or mode
30 ms typ.
Recall setting
250 ms typ.
16 k pattern transfer
600 ms typ.