Absolute Power Measurement Accuracy: The absolute power

measurement accuracy is determined by a number of factors

including calibrator accuracy, noise, sensor calibration factor

error, and the mismatch uncertainty between sensor and device

under test.

Calibrator: Provides a 50 MHz calibration signal at 51 very

accurately controlled levels from +20 to -30 dBm to dynamically

linearize the sensors.

Calibrator Frequency: 50 MHz nominal

Calibrator Connector: Type-N (f) precision connector, 50 Ω.

Settability: The 1.00 mW level in the power sweep is factory set

to ±0.7% traceable to the National Institute of Standards and

Technology (NIST).

Accuracy: ±1.2% worst case for one year, over temperature

range 15°C to 35°C.

Calibrator VSWR: < 1.05:1 (Return Loss > 33 dB)

Instrument plus Power Sensor Linearity:

Standard Sensors, CW Mode:

±0.02 dB (±0.5%) over any 20 dB range from +16 to -70 dBm

±0.02 dB + (+0 dB, -0.05 dB/dB) from +16 to +20 dBm

±0.04 dB (±1.0%) from +16 to -70 dBm

Standard Sensors, Swept Mode:

±0.03 dB (±0.7%) over any 20 dB range from +16 to -70 dBm

±0.03 dB + (+0 dB, -0.05 dB/dB) from +16 to +20 dBm

±0.06 dB (±1.4%) from +16 to -70 dBm

Low VSWR Sensors:

-64 to +20 dBm: Same as for Standard Sensors.

+20 to +30 dBm: Same as for Standard Sensors, plus an

additional ±0.13 dB (typical).

High Power Sensors:

-60 to +20 dBm: Same as for Standard Sensors.

+20 to +30 dBm: Same as for Standard Sensors, plus an

additional ±0.13 dB (typical).

True RMS Sensors, CW Mode:

±0.02 dB (±0.5%) over any 20 dB range from +20 to -30 dBm

±0.04 dB (±1.0%) from +20 to -30 dBm

True RMS Sensors, Swept Mode:

±0.03 dB (±0.7%) over any 20 dB range from +20 to -30 dBm

±0.06 dB (±1.4%) from +20 to -30 dBm

Temperature Coefficient of Linearity: < 0.3%/°C temperature

change after calibration

Zeroing Accuracy, (CW Mode, Averaging Factor = 32):

Zero set: ±50 pW

Zero drift: < ±200 pW (typical) in 1 hour at constant

temperature after at 24 hour warm-up.

Zeroing Accuracy, (Swept Mode, Averaging Factor = 32):

Zero set: ±50 pW (AC Detection), ±800 pW (DC Detection)

Zero drift: 2 nW (DC detection), typical, in 1 hour at constant

temperature after 24 hour warm-up. Zero drift not applicable

in AC detection.

Noise Uncertainty: < 50 pW, typical, at constant temperature,

measured over a 1 minute interval, two standard deviations.

Cal Factor Correction: Manual or automatic correction to power

readings to compensate for frequency response variations of the

power sensors and bridges.

Manual: Cal Factor, Cal Frequency, Off

Automatic: Sweeper

General Specifications

Temperature Range:

Operating: 0°C to 50°C, Storage: -40°C to

70°C

Power Requirements: 100/120/220/240 V ±10%, 48 to 440 Hz,

200 VA

Physical Characteristics: Dimensions: 45.1 cm (17.76 in) wide,

17.8 cm (7.00 in) high, 48.3 cm (19.00 in) deep. Weight: 16.6 Kg

(36.5 lbs)