General specifications, Electro magnetic compatibility (emc) – Atec Fluke-430 Series User Manual
Page 11
11 Fluke Corporation Fluke 430 Series Three-Phase Power Quality Analyzers
Mechanical
Size
256 mm x 169 mm x 64 mm (10 in x 6.5 in x 2.5 in)
Weight
2 kg (4.5 lb)
Power
Line power
Switchable 115 V, 230 V adapter with country specific plug
Power adapter output voltage
15 V dc to 23 V dc; use only power adapter BC430
Battery power
Rechargeable NiMH BP190 (installed)
Battery operating time
> 7 hours
Battery charging time
4 hours, 8 hours for/006 version (instrument off)
Power saving
Adjustable time for dimmed backlight with on screen power indicator
Standards
Measurement methods used
IEC 61000-4-30 Class-A
Measurement performance
Fluke 435 IEC 61000-4-30 Class-A, Fluke 434 IEC 61000-4-30 Class-B
Power quality
EN 50160
Flicker
IEC 61000-4-15
Harmonics
IEC 61000-4-7
Cross talk
Between V inputs
-60 dB @ nominal frequency
Voltage to current input
-95 dB @ nominal frequency
Safety
Compliance
IEC/EN 61010-1-2001
CAN/CSA C22.2 No 61010-1-04
UL std No 61010-1
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General requirements
Rated: 600 V CAT IV/1000 V CAT III Pollution Degree 2
Max voltage on banana input
1000 V CAT III/600 V CAT IV
Max voltage on current BNC input
42 Vpeak
Environmental
Operating temperature
0 °C to +50 °C (32 °F to 122 °F) battery only, 0 °C to +40 °C (32 °F to 104 °F) with adapter,
within spec +15 °C to +35 °C (59 °F to 95 °F)
Storage temperature
-20 °C to +60 °C (-4 °F to +140 °F)
Humidity
10 °C to 30 °C: 95 % RH non-condensing;
30 °C to 40 °C: 75 % RH non-condensing;
40 °C to 50 °C: 45 % RH non-condensing battery only
Maximum operating altitude
3000 m. Derate to 1000 V CAT II/600 V CAT III/300 V CAT IV above 2000 m
Maximum storage altitude
12 km
Warranty
Three-years on mainframe, one-year on included accessories
Printers and Interface
Type
Serial, optically isolated. Compatible with PM9080 (RS-232) or OC4USB (USB)
Baud rate
1200, 2400, 9600 to 115 kB
Print out facility (B&W only)
Via optional adapter PM9080 or PAC 91
Print protocol
Epson FX/LQ, Deskjet, LaserJet , DPU-414 or PostScript
General specifications
cont.
Electro Magnetic Compatibility (EMC)
Emission and immunity
EN61326-1
Fluke 434/435, including standard accessories, conforms with the EEC
directive 89/336 for EMC immunity, as defined by EN61326, with the
addition of the table below
Frequency
Disturbance < 0.5 %
Disturbance < 10 %
80 MHz to 400 MHz
All ranges
400 MHz to 600 MHz
All other ranges
125 V range
600 MHz to 1 GHz
All ranges
The Analyzer is susceptible for RF fields with a field strength of 10 V/m, between 400 MHz and 600 MHz (Performance criteria B).
1
Depending on clamp scaling, volt scaling 1:1
2
50 Hz/60 Hz nominal frequency according to IEC 61000-4-30
3
Add clamp accuracy and phase shift
4
The logger and mains signaling function are optional for the Fluke 434 and
standard on the Fluke 435
5
Measured on reference voltage input A/L1
6
Maximum time 9999 hours
7
Estimated duration
8
Add ±(n-1) x 2.5 ° for amplitude when using i430-flex-4pk