Measurement speed – Atec Rohde-Schwarz-ZNC User Manual

Page 5

background image

Version 03.00, September 2012

Rohde & Schwarz R&S

®

ZNC Vector Network Analyzer

5

Measurement speed

Measurement time

for 201 measurements points, with
200 MHz span, 300 kHz measurement
bandwidth with 900 MHz center frequency

< 8 ms

Measurement time per point

300 kHz measurement bandwidth,
CW mode

< 4 us

Time for measurement and data transfer

for 201 measurements points, with
800 MHz start frequency, 1 GHz stop
frequency, 300 kHz measurement
bandwidth (No additional time for data
transfer is needed, as this occurs
simultaneously during the measurement.)

typ. 8 ms

Switching time between channels

with no more than 2001 points

< 5 ms

Switching time between two preloaded
instrument settings

with no more than 2001 points

< 5 ms

Typical sweep times versus number of measurement points

2


Number of measurement points

51

201

401

1601

5001


800 MHz start frequency, 1 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth

With correction switched OFF

46 ms

176 ms

345 ms

1355 ms

4220 ms

With 2-port TOSM calibration

90 ms

350 ms

685 ms

2700 ms

8450 ms


800 MHz start frequency, 1 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth

With correction switched OFF

2.9 ms

7 ms

9 ms

23 ms

60 ms

With 2-port TOSM calibration

4.8 ms

13 ms

18 ms

46 ms

120 ms


800 MHz start frequency, 1 GHz stop frequency, AGC AUTO, 300 kHz measurement bandwidth

With correction switched OFF

2.8 ms

6 ms

8 ms

14 ms

34 ms

With 2-port TOSM calibration

4.2 ms

11 ms

14 ms

27 ms

68 ms


100 kHz start frequency, 3 GHz stop frequency, AGC LOW DIST, 1 kHz measurement bandwidth

With correction switched OFF

47 ms

177 ms

350 ms

1380 ms

4260 ms

With 2-port TOSM calibration

93 ms

352 ms

695 ms

2750 ms

8530 ms


100 kHz start frequency, 3 GHz stop frequency, AGC AUTO, 100 kHz measurement bandwidth

With correction switched OFF

4.0 ms

9 ms

14 ms

42 ms

105 ms

With 2-port TOSM calibration

7.0 ms

16 ms

25 ms

79 ms

210 ms


100 kHz start frequency, 3 GHz stop frequency, AGC AUTO, 300 kHz measurement bandwidth

With correction switched OFF

3.7 ms

8 ms

12 ms

32 ms

76 ms

With 2-port TOSM calibration

6.5 ms

14 ms

21 ms

62 ms

150 ms

2

Sweep time is to be understood as cycle time; static frequency accuracy of the instrument applies; measured with firmware version 1.20, Windows 7.