Professional leak detection - faster and better – Atec Oerlikon Leybold-PHOENIXL300 User Manual
Page 3

Professional Leak Detection - Faster and Better
Cart 300 for PHOENIXL300
Customer Benefits
PHOENIXL 300
I
Lowest detectable leak rate for
helium < 5 x 10
-12
mbar l /s
I
Faster detection times through an
averaging method which adapts
itself dynamically to the leak rates
I
Leak detection starting at an inlet
pressure of 15 mbar
I
Oil-free gas inlet system
I
Short recovery time after helium
contamination (≤ 5 min.)
I
High helium pumping speed 2.5 l/s
I
The leak detector may be moved to
a different location immediately
(≤ 30 s) after switching off
I
Conversion of the Helium leak rate
into R134a equivalent g/a, oz/yr
and ppm read out in sniffer mode
selectable
I
Integrated calibrated leak and
automatic calibration
I
3 freely selectable trigger thresholds
I
Switchable for the detection of
H
2
,
3
He and He
I
Trend/bargraph leak rate display
I
Transportation in any orientation
I
Helium background suppression
upon a single key press (Auto-Zero)
I
Measurement range of 12 decades
I
Data output and external
driving analog / RS 232
I
Documentation and qualification of
the leak rates acc. to ISO 9000
PHOENIXL 300 Dry
I
Entirely oil-free vacuum system
I
Purge function for preventing
helium accumulations in the
system
PHOENIXL 300 Modul
I
Additional forevacuum system for
testing of large volume vessels
PHOENIXL 300 Cart-Option
I
Modular and mobile system;
The PHOENIXL 300 is configured
according to customer specificati-
ons depending on the demanded
leak rate, timings and type of gas.
Design Characteristics
I
Newly designed operating panel
and graphic display for easy and
convenient operation
I
Angled operating panel for easy
visibility of the instrument status
I
Innovative oil-free gas inlet system
I
Maintenance friendly concept –
the mechanical connections have
been separated from the electrical
connections and have each been
arranged in a protected connection
strip
I
Oil level glass at the housing for
simple oil level checks
I
Extremely long service life of the
cathodes due to their
Iridium/Yttrium coating
I
Distinctive, modern design
I
The optional remote control unit is
equally suited for right handers and
left handers
Additional Benefits
The Oerlikon Leybold Vacuum
PHOENIXL300 helium leak detectors
are so-called counterflow leak detec-
tors which contain many well proven
own components, like for example:
I
TRIVAC rotary vane pumps for
rapid evacuation of the device
under test and for producing a
high gas flow in the sniffer mode
I
Turbomolecular pumps TW 70 for
improving the pumping speed
and the helium background
characteristic
I
A 180° mass spectrometer with
an excellent track record already
in the predecessors to the
PHOENIXL300 leak detector
I
The optimised valve block
ensures, compared to the design
used in the predecessors, a
significantly reduced recovery time
after suffering helium contamination
and a high pumping speed at the
inlet