Expandable Test Systems
IEC / EN 61000-4-5 EDITION 2 & UPCOMING EDITION 3 SURGE COMBINATION WAVE
Output voltage
0.2 – 5.0 kV ±10%
Output current
0.1 – 2.5 kA ±10%
Voltage rise time
1.2 µs ±30%
Current rise time 8 µs ±20%
Voltage duration
50 µs ±20%
Current duration 20 µs ±20%
Polarity
postive / negative /
alternate
Integrated single
phase CDN
264 V AC / 16 A
220 V DC / 10 A
Output impedance 2 Ohm
Phase sync
0 – 359° with 1° steps or
asynchronous mode
Impulse trigger
automatic 2 s – 100 min
manual
external trigger input
Counter preselect 1 – 1'000 / infinite
Counter
100'000
Peak voltage
monitor
BNC output: 1000:1
display: 3 digits
Peak current
monitor
BNC output: 1 kA/V
display: 3 digits
IEC / EN 61000-4-11 EDITION 2 AND IEC / EN 61000-4-29 DIPS & INTERRUPTS
Max. voltage
264 V AC/DC
Interrupt time
0.5 period – 800 peri-
ods
100 µs – 1000 minutes
Max. current
16 A AC/DC continuous
20 A for 5 s
23 A for 3 s
40 A for 3 s
> 500 A inrush current
Interval time
1 period – 800 periods
synch
100 µs – 1000 minutes
asynch
Trigger
automatic
manual
external trigger input
Test time
1 s – 1000 minutes
infinite
Interrupt
dip level
0%
0% – 99% with external
voltage source
Phase sync
0 – 359° 16 / 40 / 50 /
60 Hz
asynchronous mode
RMS voltage
monitor
BNC output: 100:1
display: 4 digits
RMS current
monitor
BNC output: 10 A/V
display: 4 digits