Technical data - axos – Atec Haefely-AXOS Series User Manual
Page 14
![background image](/manuals/319195/14/background.png)
14
AXOS SERIES
GENERAL DATA
Control power
85 V - 264 V
50/60 Hz
Dimensions
(W x H x D)
22“ / 6U
(45 x 27 x 50 cm)
User test storage
unlimited
Weight
30 kg
Remote interface
Ethernet, RJ45
USB
for USB memory stick
Display
7“ / 800x480 / 24 bit
with touch-screen
AUX. interface
D-sub 37p for external
CDN, external trans-
former etc.
External trigger
input
5 V TTL
Synch input
BNC, 10 V – 264 V AC
Trigger output
5 V TTL
External start / stop
input
5 V TTL, starts / stops
predefined test se-
quence
EUT failed input
5 V TTL
Analog output
0 – 10 V, for use with
external options
Warning lamp
output
2 x 24 V / 1 A DC
Safety circuit
stops the test when
unlocked
IEC / EN 61000-4-4 EDITION 2 & 3 EFT / BURST
Output voltage
0.2 – 5.0 kV ±10% at
coaxial output
Spike frequency
1 Hz – 1 MHz
Polarity
pos / neg / alternate
Burst duration
10 us – 1 s
Output impedance
50 Ohms
Burst period
1 ms – 10 s
Rise time
5 ns ±30%
Test time
1 s– 1000 minutes
Impulse duration
50 ns ±30% at 50
Ohm
50 ns –15 +100 ns at
1000 Ohm
Trigger
automatic, manual,
external trigger input
Burst mode
normal, continuous,
real, random
Integrated single
phase coupling /
decoupling network
264 V AC / 16 A
220 V DC / 10 A
IEC / EN 61000-4-5 EDITION 2 & UPCOMING EDITION 3 SURGE COMBINATION WAVE
Output voltage
0.2 – 7.0 kV ±10%
Output current
0.1 – 3.5 kA ±10%
Voltage rise time
1.2 us ±30%
Current rise time
8 us ±20%
Voltage duration
50 us ±20%
Current duration
20 us ±20%
Polarity
pos / neg / alternate
Integrated single
phase CDN
264 V AC / 16 A
220 V DC / 10 A
Output impedance
2 Ohms
Phase sync
0 – 359° with 1° steps
or asynchronous
mode
Impulse trigger
automatic 2 s – 100 min
manual
external trigger input
Counter preselect
1 – 1000 / infinite
Counter
100000
Peak voltage
monitor
BNC output: 1000:1
display: 3 digits
Peak current
monitor
BNC output: 1 kA/V
display: 3 digits
TECHNICAL DATA - AXOS