Model PRO-HPULSE

Pulse Magnetic Field for compliant testing per IEC 61000-4-9 and EN 61000-4-9

Field Pulse

8/20µs

Field Amplitude

100A/m - 1000A/m, ±10%

Resolution

5A/m

Coil Factor

0.65 to 1.00

Options Model

CM-HMON:

Measurement probe for power

frequency magnetic fields
Model CM-HCOIL: 1m x 1m magnetic field coil

Model PRO-PQF

Dips and Interrupts for compliant testing IEC 61000-4-11 Edition 2, and EN 61000-4-11

Dips

40%, 70%, 80%

Interrupts

0% (short and open)

Transition Time

1µs - 5µs

Inrush

Minimum 250Amps @ 100 - 120V,
Minimum 500Amps @ 220 - 240V

AC Voltage

50 - 250VAC, 50/60Hz

AC Current

16A max.**

PQF Sync Output

5V signal occurs at each dip or interrupt transition

Options

Model PQF-QUAL: Circuit per IEC 61000-4-11 for
testing PQF generator inrush capability

PQF Waveform Monitoring

Voltage Input Connection

Fixed, L1 to L2

Voltage Attenuation

100:1 ±5%

Current Input Connection

Fixed, L1

Peak Current

Minimum 500A inrush into 1700µF

Current Attenuation

200:1 ±5%

OPTIONAL COUPLERS/DECOUPLERS

Model CM-3CD-16 & CM-3CD-32

*

Semi-automatic, stand alone, three-phase AC/DC mains coupler/decouplers for EFT &
Surge per IEC 61000-4-4, Edition 2 and IEC 61000-4-5

ELECTRICAL

Waveforms

EFT: 5/50ns, per IEC 61000-4-4
Surge: Combination wave: 1.2/50µs open-circuit
voltage, 8/20µs short-circuit current, per IEC 61000-4-5

Maximum Surge

6.6kV, 3.3kA

Voltage & Current

Maximum EFT Voltage

4.4kV

Coupling Modes

EFT: L1, L2, L3, N or PE
Surge Hi: L1, L2, L3 or N
Surge Lo: L1, L2, L3, N or PE

*

Not available for delivery until October 2004

COUPLER/DECOUPLERS

AC Voltage

50 to 250V, 50/60Hz line to ground, 50 to 433V
line to line

AC Current

CM-3CD-16: 16A/phase continuous
CM-3CD-32: 32A/phase continuous

DC Current

CM-3CD-16: 16A up to 48V

8A up to 110V
1.2A up to 220V
0.3A up to 440V

CM-3CD-32: 25A up to 48V

8A up to 220V
1.2A up to 220V
0.3A up to 440V

EUT Mains

Safety Sockets

Output Connectors

POWER REQUIREMENTS

Input Voltage

90-250VAC, 50/60Hz

Input Current

1A at 120VAC; 0.5A at 240VAC

Model CM-I/OCD

I/O coupler/decoupler - provides the ability to couple surges from EMCPro PLUS or any
surge simulator, to I/O or data lines per IEC 61000-4-5

ELECTRICAL

Waveforms

Designed to couple combination waves of 1.2/50µs
open-circuit voltage, 8/20µs short-circuit current
supplied by option PRO-SURGE with the KeyTek
EMCPro PLUS

Repetition Rate

Up to 5 per minute at 4.4kV

Data Line Frequency

To greater than 100kHz without significant degradation
when CM-I/OCD-HS is installed. Option CM-I/OCD-HS
is recommended for data line frequencies greater than
1kHz

Number of Lines

Eight lines - any line can be surged to any other line or
ground

Maximum Surge Voltage

4.4kV

Maximum Signal Line Voltage

200V

Maximum Signal Line Current

1A AC or DC

Clamping

Selectable built-in clamps of 20V and 220V; external
bias input for other clamp levels

Available Options

CM-I/OCD-HS: Internally-installed option provides
selectable parallel resistors (400s, 200s, 100s) - highly
recommended for data line frequencies greater than
1kHz.

Model CM-TELCD

Telecom line coupler/decoupler - provides the ability to couple both the telecom wave and
combination wave per IEC 61000-4-5

ELECTRICAL

Waveforms

Designed to couple 1.2/50µs combination or 10/700µs
telecom waves

Telecom Line Frequency

To 100kHz without significant degradation

Number of Lines

Up to four lines - one or two pairs of balanced Telecom
lines

Maximum Surge Voltage

4.4kV

Maximum Signal Line Voltage

200V

Maximum Signal Line Current

1A AC or DC

Clamping

Selectable built-in clamps of 20V and 225V: external
bias input for other clamp levels