Model pro-hpulse, Model pro-pqf, Pqf waveform monitoring – Atec Keytek-EMCPro-PLUS User Manual
Page 3: Model cm-i/ocd, Model cm-telcd
![background image](/manuals/319183/3/background.png)
Model PRO-HPULSE
Pulse Magnetic Field for compliant testing per IEC 61000-4-9 and EN 61000-4-9
Field Pulse
8/20µs
Field Amplitude
100A/m - 1000A/m, ±10%
Resolution
5A/m
Coil Factor
0.65 to 1.00
Options Model
CM-HMON:
Measurement probe for power
frequency magnetic fields
Model CM-HCOIL: 1m x 1m magnetic field coil
Model PRO-PQF
Dips and Interrupts for compliant testing IEC 61000-4-11 Edition 2, and EN 61000-4-11
Dips
40%, 70%, 80%
Interrupts
0% (short and open)
Transition Time
1µs - 5µs
Inrush
Minimum 250Amps @ 100 - 120V,
Minimum 500Amps @ 220 - 240V
AC Voltage
50 - 250VAC, 50/60Hz
AC Current
16A max.**
PQF Sync Output
5V signal occurs at each dip or interrupt transition
Options
Model PQF-QUAL: Circuit per IEC 61000-4-11 for
testing PQF generator inrush capability
PQF Waveform Monitoring
Voltage Input Connection
Fixed, L1 to L2
Voltage Attenuation
100:1 ±5%
Current Input Connection
Fixed, L1
Peak Current
Minimum 500A inrush into 1700µF
Current Attenuation
200:1 ±5%
OPTIONAL COUPLERS/DECOUPLERS
Model CM-3CD-16 & CM-3CD-32
*
Semi-automatic, stand alone, three-phase AC/DC mains coupler/decouplers for EFT &
Surge per IEC 61000-4-4, Edition 2 and IEC 61000-4-5
ELECTRICAL
Waveforms
EFT: 5/50ns, per IEC 61000-4-4
Surge: Combination wave: 1.2/50µs open-circuit
voltage, 8/20µs short-circuit current, per IEC 61000-4-5
Maximum Surge
6.6kV, 3.3kA
Voltage & Current
Maximum EFT Voltage
4.4kV
Coupling Modes
EFT: L1, L2, L3, N or PE
Surge Hi: L1, L2, L3 or N
Surge Lo: L1, L2, L3, N or PE
*
Not available for delivery until October 2004
COUPLER/DECOUPLERS
AC Voltage
50 to 250V, 50/60Hz line to ground, 50 to 433V
line to line
AC Current
CM-3CD-16: 16A/phase continuous
CM-3CD-32: 32A/phase continuous
DC Current
CM-3CD-16: 16A up to 48V
8A up to 110V
1.2A up to 220V
0.3A up to 440V
CM-3CD-32: 25A up to 48V
8A up to 220V
1.2A up to 220V
0.3A up to 440V
EUT Mains
Safety Sockets
Output Connectors
POWER REQUIREMENTS
Input Voltage
90-250VAC, 50/60Hz
Input Current
1A at 120VAC; 0.5A at 240VAC
Model CM-I/OCD
I/O coupler/decoupler - provides the ability to couple surges from EMCPro PLUS or any
surge simulator, to I/O or data lines per IEC 61000-4-5
ELECTRICAL
Waveforms
Designed to couple combination waves of 1.2/50µs
open-circuit voltage, 8/20µs short-circuit current
supplied by option PRO-SURGE with the KeyTek
EMCPro PLUS
Repetition Rate
Up to 5 per minute at 4.4kV
Data Line Frequency
To greater than 100kHz without significant degradation
when CM-I/OCD-HS is installed. Option CM-I/OCD-HS
is recommended for data line frequencies greater than
1kHz
Number of Lines
Eight lines - any line can be surged to any other line or
ground
Maximum Surge Voltage
4.4kV
Maximum Signal Line Voltage
200V
Maximum Signal Line Current
1A AC or DC
Clamping
Selectable built-in clamps of 20V and 220V; external
bias input for other clamp levels
Available Options
CM-I/OCD-HS: Internally-installed option provides
selectable parallel resistors (400s, 200s, 100s) - highly
recommended for data line frequencies greater than
1kHz.
Model CM-TELCD
Telecom line coupler/decoupler - provides the ability to couple both the telecom wave and
combination wave per IEC 61000-4-5
ELECTRICAL
Waveforms
Designed to couple 1.2/50µs combination or 10/700µs
telecom waves
Telecom Line Frequency
To 100kHz without significant degradation
Number of Lines
Up to four lines - one or two pairs of balanced Telecom
lines
Maximum Surge Voltage
4.4kV
Maximum Signal Line Voltage
200V
Maximum Signal Line Current
1A AC or DC
Clamping
Selectable built-in clamps of 20V and 225V: external
bias input for other clamp levels