Model PRO-ESD
ESD per IEC 61000-4-2 and EN 61000-4-2
Trigger Modes
One shot manual, multi-shot tripod
Repetition Rate
Single shot, 1pps or 20pps
Air Discharge Voltage
500V - 8.8kV ±10%
Contact Discharge Voltage
500V - 4.4kV ±10%
Discharge Capacitor
150pF ±10%
Discharge Resistance
330
Ω
±10%
Charging Resistance
50M
Ω
- 100M
Ω
Polarity
Front panel or software controlled
Shot Counter
1 - 999 discharges
Energy Storage
5.8mJ @ 8.8kV
Model PRO-EFT
EFT per IEC 61000-4-4 Edition 2, EN 61000-4-4 and ANSI C62.41
Voltage Waveform
5/50ns ±30%
Peak Voltage
250V - 4.4kV ±5%
Burst Period
300ms ±10%
Burst Duration
15ms ±20%, for pulse frequencies uo to 5kHz, 0.75ms
above 5kHz
Frequency
1-100kHz, in 0.5kHz steps, ±10%
DC Blocking Capacitor
10nF (internal)
Options
Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT &
surge coupler/decoupler
Model CM-CCL: Capacitive coupling clamp
Model CM-CCLC: Coupling clamp cover
Model EFT-ATTN: EFT attenuator for oscilloscope
monitoring
Model PRO-SURGE
Surge for compliant testing per IEC 61000-4-5, EN 61000-4-5, ANSI C62.41 Category B and
UL 1449
Voltage Waveform
1.2/50µs
Peak Voltage
250V - 6.6kV ±5%, 12
Ω
mode
250V - 6.0kV ±5%, 2
Ω
mode
Peak Current
125A - 3.3kA ±10%
Additional 10
Ω
Resistor
Software selectable
Repetition Rate
Up to 4 per minute
Open-circuit Voltage
Front time: 1.2µs ±30%
Duration: 50 µs ±20%
1
Undershoot:
≤
30%
Short-circuit Current
Front time: 8.0µs ±20%
Duration
*
50µs ±20%
Undershoot
≤
30%
Line sync accuracy
±15%, 50 - 277VAC
Options Model
CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT
& surge coupler/decoupler
Model CM-I/OCD: External 8 line coupler/decoupler
for I/O signal lines
Model CM-I/OCD-HS: High speed I/OCD option for
testing data rates to >100kHz
*
Durations are reduced in 12
Ω
mode and when coupling multiple lines to PE
Model PRO-RING
**
Ring Wave Surge per ANSI C62.41 Cat. A, B, and UL 864
Voltage Waveform
100kHz damped cosine
Peak Voltage
250 - 6.6kV ±5%
Repetition Rate
<4/minute at 6kV, faster at lower voltages
Open-circuit Voltage
Rise Time: 0.5µs ±30%
Short-circuit Current
Vp/Ip: 12
Ω
± 3
Ω
or 30
Ω
± 8
Ω
software selectable
Options Model
CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT
& surge coupler/decoupler
Model PRO-TELECOM
**
Surge Telecom compliant testing per IEC 61000-4-5, EN 61000-4-5, FCC Part 68, ITU K.17,
K.20, K.21 and ETSI
Voltage Waveform
10/700µs (9/720µs FCC Part 68)
Peak Voltage
250V - 6.6kV ±5%
Peak Current
6.25 - 165A +10/-0%, 40
Ω
mode
Repetition Rate
Up to 4 per minute
Open-circuit Voltage
Front time: 7.0µs to 11.7µs
Duration: 576µs to 840µs
Short-circuit Current
Front time: 3.5µs to 6.5µs
Duration: 256µs to 384µs
Options
Model CM-TELCD: External coupler for telecom lines
Surge Waveform Monitoring
Lines Monitored
Monitors are automatically switched to match
generator coupling mode
Open-circuit Voltage
1000:1 ±10%
Short-circuit
200:1 ±7%
Current Attenuation
Model PRO-HPOWER
Power Frequency Magnetic Field for compliant testing per IEC 61000-4-8 and EN 61000-4-8
Field Frequency
50Hz/60Hz
Field Amplitude
0.5 - 4A/m, in 0.25A steps, ±10% (with CM-HCOIL)
up to 100A/m with optional external HPOWER-EXT
AC Source
Internal
Resolution
0.25A minimum
Coil Factor
0.65 to 1.00
Coil Resistance
0.05
Ω
maximum
Options Model
CM-HMON: Measurement probe for power
frequency magnetic fields
Model CM-HCOIL: 1m x1m magnetic field coil
Model HPOWER - EXT: External generator for power
frequency magnetic field to 30A/m
**
PRO-TELECOM and PRO-RING can not be installed in same unit.