beautypg.com

Characteristic values, Reference conditions, Power supply – Atec Gossen-Metrawatt-5000DPI User Manual

Page 2: Power supply profitest 204hp/hv, Ambient conditions

background image

METRISO 5000 D-PI
Digital High-Voltage Insulation Tester

2

GMC-I Messtechnik GmbH

Characteristic Values

Measuring Ranges:

Standard

DIN EN 61557-1:2007
DIN EN 61557-2:2008

VDE Regulation

VDE 0413 Part 1:2007
VDE 0413 Part 2:2008

Insulation Resistance

Test duration: automatic (until measured value is stable),

manual (1 to 120 s) or continuous measurement (lock function)

Polarization Index (PI), Absorption Ratio (DAR)

PI and DAR are calculated values. The specifications of the insulation measurement
are applicable.

Insulation Test Voltage

Variable test voltages are adjustable in increments of 50 V
Short-circuit current up to 1.00 kV, test voltage

 2 mA

Voltage Measurement

Frequency Measurement

Voltage of measuring quantity: 50 V ... 1 kV

Breakdown Voltage

Capacitance Measurement

Dielectric Discharge (DD)

Reference Conditions

Ambient
temperature +23

C 2 K

Relative humidity

40 ... 60%

Measured quantity
frequency

50 Hz

10 Hz (during voltage

measurement)

Line voltage
waveshape

Sinusoidal, deviation between RMS and
rectified value < 1%

Power Supply

Line voltage

207 V

253 V / 49 Hz

61 Hz

or (depending on country-specific version)
108 V

132 V / 59 Hz

61 Hz

Power consumption

< 18 VA

Storage batteries

NiMH 9.6 V, 3 Ah, charging period 6 hours

Number of
measurements
at nominal current
as per VDE 0413

700

Power Supply PROFITEST 204HP/HV

Line voltage

207 V

253 V / 49 Hz

61 Hz

Power consumption

PROFITEST 204HP/2.5kV: max. 700 VA
PROFITEST 204HV/5.4kV: max. 100 VA

Ambient Conditions

Accuracy

0

C ... + 40 C

Operating temperature

5 C ... + 40 C

Storage temperature

20 C ... + 60 C (without batteries)

Relative humidity

max. 75%, no condensation allowed

Elevation

to 2000 m

Deployment

indoors, outdoors: only in the specified
abient conditions

Display
Range [

]

Measuring
Range

Test Voltage

Intrinsic
Uncertainty

Measuring
Uncertainty

0.00 M ... 50.0 G 0.60 M ...10.0 G

100 V ... 250 V

±(7% rdg. + 6d)

±(10% rdg. + 8 d)

>10.0 G ... 50.0 G

±(7% rdg. + 6d)

±(10% rdg. + 8 d)

0.00 M ... 250 G 0.40 M ... 50.0 G

> 250 V ... 1.00 kV ±(7% rdg. + 6d)

±(10% rdg. + 8 d)

>50.0 G ... 250 G

±(7% rdg. + 6d)

±(10% rdg. + 8 d)

0.00 M ... 999 G 0.40 M ... 200 G

>1.00 kV ... 5.00kV ±(7% rdg. + 6d)

±(10% rdg. + 8 d)

>200 G ... 999 G

±(7% rdg. + 6d)

±(10% rdg. + 8 d)

t1 [min]

t2 [min]

Limit [min]

PI

00:00 ... 01:00 ... 99:50

00:00 ... 10:00 ... 99:50

0.10 ... 4.00 ... 9.80

DAR

00:00 ... 00:30 ... 99:50

00:00 ... 01:00 ... 99:50

0.10 ... 1.60 ... 9.80

Nominal Values
of Test Voltage

Variable
Test Voltage

Nominal
Current

Intrinsic
Uncertainty

100 V, 250 V, 500 V, 1.00 kV

 1.0 mA

0 ... +25% rdg.

1.50 kV, 2.00 kV, 2.50 kV

 0.4 mA

± 5% rdg.

5.00 kV

 0.1 mA

± 3.5% rdg.

100 V...1.00 kV

 1.0 mA

± 15% rdg.

> 1.00 kV...2.50 kV

 0.4 mA

± 5% rdg.

> 2.50 kV...5.00 kV

 0.1 mA

± 3.5% rdg.

Measuring range

Frequency
[Hz]

Impe-
dance

Intrinsic
Uncertainty

Measuring
Uncertainty

test voltage dc
50 V ... 5.00 kV

(2.5% rdg. + 5 d)

(5% rdg. + 5 d)

50 V ... 1.00 kV ac/dc

15 ... 500

1 M

(2.5% rdg. + 2 d)

(5% rdg. + 5 d)

50 V ... 1.00 kV ac/dc

>500...1 k

1 M

(10% rdg. + 2 d)

(12.5% rdg. + 5 d)

Measuring Range

Impedance

Intrinsic
Uncertainty

Measuring
Uncertainty

15.0 Hz

 1.00 kHz

1 M

(0.5% rdg. + 2 d) (1% rdg. + 2 d)

Parameters

Setting Range

Intrinsic
Uncertainty

Measuring
Uncertainty

Voltage range

100

5000 V

(

10% rdg. + 8 d)

(

15% rdg. + 10 d)

Rise time

5

300 s

Measuring time

1

120 s / auto /

cont. measurement

Display
Range

Measuring
Range

Test
Voltage

Intrinsic
Uncertainty

Measuring
Uncertainty

0.00

10.0 F 0.10 5.00 F

100

450 V 

(

10% rdg. + 5 d)

(

15% rdg. + 8 d)

500

5 kV

(

5% rdg. + 5 d)

(

10% rdg. + 8 d)

Limit

DD

0.10 ... 2.00 ... 9.80