Characteristic values, Reference conditions, Power supply – Atec Gossen-Metrawatt-5000DPI User Manual
Page 2: Power supply profitest 204hp/hv, Ambient conditions
METRISO 5000 D-PI
Digital High-Voltage Insulation Tester
2
GMC-I Messtechnik GmbH
Characteristic Values
Measuring Ranges:
Standard
DIN EN 61557-1:2007
DIN EN 61557-2:2008
VDE Regulation
VDE 0413 Part 1:2007
VDE 0413 Part 2:2008
Insulation Resistance
Test duration: automatic (until measured value is stable),
manual (1 to 120 s) or continuous measurement (lock function)
Polarization Index (PI), Absorption Ratio (DAR)
PI and DAR are calculated values. The specifications of the insulation measurement
are applicable.
Insulation Test Voltage
Variable test voltages are adjustable in increments of 50 V
Short-circuit current up to 1.00 kV, test voltage
2 mA
Voltage Measurement
Frequency Measurement
Voltage of measuring quantity: 50 V ... 1 kV
Breakdown Voltage
Capacitance Measurement
Dielectric Discharge (DD)
Reference Conditions
Ambient
temperature +23
C 2 K
Relative humidity
40 ... 60%
Measured quantity
frequency
50 Hz
10 Hz (during voltage
measurement)
Line voltage
waveshape
Sinusoidal, deviation between RMS and
rectified value < 1%
Power Supply
Line voltage
207 V
253 V / 49 Hz
61 Hz
or (depending on country-specific version)
108 V
132 V / 59 Hz
61 Hz
Power consumption
< 18 VA
Storage batteries
NiMH 9.6 V, 3 Ah, charging period 6 hours
Number of
measurements
at nominal current
as per VDE 0413
700
Power Supply PROFITEST 204HP/HV
Line voltage
207 V
253 V / 49 Hz
61 Hz
Power consumption
PROFITEST 204HP/2.5kV: max. 700 VA
PROFITEST 204HV/5.4kV: max. 100 VA
Ambient Conditions
Accuracy
0
C ... + 40 C
Operating temperature
5 C ... + 40 C
Storage temperature
20 C ... + 60 C (without batteries)
Relative humidity
max. 75%, no condensation allowed
Elevation
to 2000 m
Deployment
indoors, outdoors: only in the specified
abient conditions
Display
Range [
]
Measuring
Range
Test Voltage
Intrinsic
Uncertainty
Measuring
Uncertainty
0.00 M ... 50.0 G 0.60 M ...10.0 G
100 V ... 250 V
±(7% rdg. + 6d)
±(10% rdg. + 8 d)
>10.0 G ... 50.0 G
±(7% rdg. + 6d)
±(10% rdg. + 8 d)
0.00 M ... 250 G 0.40 M ... 50.0 G
> 250 V ... 1.00 kV ±(7% rdg. + 6d)
±(10% rdg. + 8 d)
>50.0 G ... 250 G
±(7% rdg. + 6d)
±(10% rdg. + 8 d)
0.00 M ... 999 G 0.40 M ... 200 G
>1.00 kV ... 5.00kV ±(7% rdg. + 6d)
±(10% rdg. + 8 d)
>200 G ... 999 G
±(7% rdg. + 6d)
±(10% rdg. + 8 d)
t1 [min]
t2 [min]
Limit [min]
PI
00:00 ... 01:00 ... 99:50
00:00 ... 10:00 ... 99:50
0.10 ... 4.00 ... 9.80
DAR
00:00 ... 00:30 ... 99:50
00:00 ... 01:00 ... 99:50
0.10 ... 1.60 ... 9.80
Nominal Values
of Test Voltage
Variable
Test Voltage
Nominal
Current
Intrinsic
Uncertainty
100 V, 250 V, 500 V, 1.00 kV
1.0 mA
0 ... +25% rdg.
1.50 kV, 2.00 kV, 2.50 kV
0.4 mA
± 5% rdg.
5.00 kV
0.1 mA
± 3.5% rdg.
100 V...1.00 kV
1.0 mA
± 15% rdg.
> 1.00 kV...2.50 kV
0.4 mA
± 5% rdg.
> 2.50 kV...5.00 kV
0.1 mA
± 3.5% rdg.
Measuring range
Frequency
[Hz]
Impe-
dance
Intrinsic
Uncertainty
Measuring
Uncertainty
test voltage dc
50 V ... 5.00 kV
—
—
(2.5% rdg. + 5 d)
(5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc
15 ... 500
1 M
(2.5% rdg. + 2 d)
(5% rdg. + 5 d)
50 V ... 1.00 kV ac/dc
>500...1 k
1 M
(10% rdg. + 2 d)
(12.5% rdg. + 5 d)
Measuring Range
Impedance
Intrinsic
Uncertainty
Measuring
Uncertainty
15.0 Hz
1.00 kHz
1 M
(0.5% rdg. + 2 d) (1% rdg. + 2 d)
Parameters
Setting Range
Intrinsic
Uncertainty
Measuring
Uncertainty
Voltage range
100
5000 V
(
10% rdg. + 8 d)
(
15% rdg. + 10 d)
Rise time
5
300 s
—
—
Measuring time
1
120 s / auto /
cont. measurement
—
—
Display
Range
Measuring
Range
Test
Voltage
Intrinsic
Uncertainty
Measuring
Uncertainty
0.00
10.0 F 0.10 5.00 F
100
450 V
(
10% rdg. + 5 d)
(
15% rdg. + 8 d)
500
5 kV
(
5% rdg. + 5 d)
(
10% rdg. + 8 d)
Limit
DD
0.10 ... 2.00 ... 9.80