Options, For electrical safety testers, Specifications – Atec Kikusui-TOS6100 User Manual
Page 3: Outlines
Page3
Copyright © 2002 KIKUSUI ELECTRONICS CORP. All rights reserved.
http://www.kikusui.co.jp/
149-10A
Specifications
Operating System
Double integration system (sampling
cycle: 3 times/sec)
DC Voltage
M e a s u r i n g r a n g e : 0 . 5 0 0 k V t o
10,000kV
Accuracy:
±
(0.5% of reading + 0.03%
of range)
Input resistance: 1000 M
Ω
±
2%
AC Voltage
M e a s u r i n g r a n g e : 0 . 5 0 0 k V t o
10,000kV
Accuracy:
±
(1% of reading + 0.05% of
range)
Frequency characteristics: 50/60 Hz
(sine wave rms value display of mean
value response)
Input resistance: 1000 M
Ω
±
2%
Power Requirements
1 0 0 V
±
1 0 % ,
approx. 10 VA
Dimensions (MAX)
134W
×
164H
×
270D mm
(140W
×
189H
×
350D mm)
Weight
approx. 3 kg
Accessories
TL05-TOS high-voltage test lead: 1
HTL-2.5DH high-voltage coaxial
cable: 1
Outlines
■Model 149-10A High-Voltage Digital Voltmeter:
Accommodation of Output Voltage Calibration of
Withstanding Voltage and
Insulation Resistance Testers
❏Measurement of high voltages (AC/DC) of up to 10 kV
maximum.
❏Large 4
1
/
2
digit LED display
❏High measuring accuracy and input resistance
❏Light weight of only 3.2 kg
❏Compact design
❏Excellent ease of maintenance
OPTIONS
FOR ELECTRICAL SAFETY TESTERS
■ Model RL01-TOS UL Resistance Load: Checking of Output
Voltage of Withstanding Voltage Testers
This device is described in section 125, paragraph 2-1B1 of
UL1492. The RL01-TOS is a variable load resistor for checking
the output voltage of withstanding voltage testers used in dielectric
strength testing on production lines. (Complies with UL regulations
including UL1270, UL1409 and UL1410.)
Specifications
Resistors:
120, 159, 210, 279, 369, 489, 648, 858,
1,137, 1,500, 1,989 and 2,148 k
Ω
Resistance Accuracy
+1%,-0% of nominal value when set
to 120 k
Ω
,
±
1% of nominal value when
set to other values
Maximum OperatingVoltage
1300 V (continuous rating)
Maximum Overload Voltage
1400 V for 5 seconds (application may
not be repeated within 1 minute)
Dimensions (MAX)
200W
×
100H
×
260D mm
(210W
×
120H
×
295D mm)
Weight
approx. 2.6 kg
Accessories
TL04-TOS high-voltage test lead: 2
TL05-TOS high-voltage test lead: 1
RL01-TOS
Rear Panel