Options, For electrical safety testers, Outlines – Atec Kikusui-TOS6100 User Manual
Page 2: Specifications
![background image](/manuals/319089/2/background.png)
Page2
Copyright © 2002 KIKUSUI ELECTRONICS CORP. All rights reserved.
http://www.kikusui.co.jp/
OPTIONS
FOR ELECTRICAL SAFETY TESTERS
Outlines
■Model TOS1200 Withstanding Voltage Tester Current
Calibrator: For calibration of current sensitivity of
withstanding voltage testers
❏Calibration of Leakage Current Detection Sensitivity
As a result of containing an internal test load resistor, the
TOS1200 enables switching of ammeter sensitivity and load
resistance values automatically for easy calibration of the
sensitivity of leakage current detection.
❏Direct Reading of Error from Error Display Scale
The main scale consists of an error display scale ranging
from -10% to +10% around a central full scale value. Thus,
if the leakage current detection sensitivity of the withstanding
voltage tester is a value of 1, 2 and 5 steps or 1, 2, 4 and 8
steps, the error can be read directly from the scale. The
auxiliary scale consists of a ratio display scale in which the
full scale value is taken to be 1.1, thereby enabling
accommodation of values outside the range of the main scale.
❏Ammeter Ranges
The ranges of the ammeter consist of 0.5, 1, 2, 5, 10, 20, 50
and 100 mA, and values equal to 0.8 times those values can
be set. This is convenient for withstanding voltage testers
of the type in which leakage current detection sensitivity is
set digitally (1, 2, 4 and 8 steps).
❏Eliminates Need for Power Supply
The TOS1200 offers outstanding portability since it does
not require a power supply.
❏AC/DC Selection Switch
Allowed Input Time
0.5/1/2/5 mA ranges: Continuous
10/20/50/100 mA ranges: 60 sec.
Max. 1/3 of duty cycle
Dimensions (MAX)
134W
×
164H
×
270D mm
(140W
×
189H
×
320D mm)
Weight
approx. 3.5 kg
Accessories
TL04-TOS high-voltage test lead: 1
TOS1200
Resistance[k
Ω
]
Range[mA]
2000
1000
500
200
Specifications
Measuring Function
Measurement of current values and
error (%) for AC (50/60 Hz) and DC at
a test voltage of 1000 V
Measuring Ranges
8 ranges consisting of 0.5/1/2/5/10/20/
50/100 mA along with values equal to
0.8 times the values of those ranges
(for 1, 2, 4 and 8 steps)
Ammeter Scale
Main scale: Direct-reading error
display scale over a range of
±
10% of
the above full scale values
Auxiliary scale: Ratio scale of 0 to 1.1
times the above full scale values
(equivalent to 0% display of main scale
when the ratio is equal to 1)
Ammeter Accuracy
Main scale:
±
1% of reading
Auxiliary scale:
±
3% of full scale value
Ammeter Indication
D C / A C ( s i n e w a v e r m s v a l u e
calibration of mean value response)
Load Resistance
0.5
1
2
5
Resistance[k
Ω
]
Range[mA]
10
20
50
100
100
50
20
10