Specifications torkel 820, Torkel 820 – Atec Megger-TORKEL-820 User Manual
Page 3
TORKEL 820
Battery Load Unit
Specifications TORKEL 820
Specifications are valid at nominal input voltage and an ambient
temperature of +25°C, (77°F). Specifications are subject to change
without notice.
Environment
Application field
The instrument is intended for use in
high-voltage substations and industrial
environments.
Temperature
Operating
0°C to +40°C (32°F to +104°F)
Storage & transport -40°C to +70°C (-40°F to +158°F)
Humidity
5% – 95% RH, non-condensing
CE-marking
LVD
2006/95/EC
EMC
2004/108/EC
General
Mains voltage
100 – 240 V AC, 50 / 60 Hz
Power consumption
150 W (max)
Protection
Thermal cut-outs, automatic overload
protection
Dimensions
Instrument
210 x 353 x 700 mm
(8.3” x 13.9” x 27.6”)
Transport case
265 x 460 x 750 mm
(10.4” x 18.1” x 29.5”)
Weight
22.3 kg (49.2 lbs)
40.4 kg (89.1 lbs) with accessories and
transport case
Display
LCD
Available languages
English, French, German, Spanish, Swe-
dish
Measurement section
Current measurement
Display range
0.0 – 2999 A
Basic inaccuracy
±(0.5% of reading +0.2 A)
Resolution
0.1 A
Internal current measurement
Range
0 – 270 A
Input for clamp-on ammeter
Range
0 – 1 V
mV/A-ratio
Software settable, 0.3 to 19.9 mV/A
Input impedance
>1 MΩ
Voltage measurement
Display range 0.0 – 60 V
Basic inaccuracy
±(0.5% of reading +0.1 V)
Resolution
0.1 V
Time measurement
Basic inaccuracy
±0.1% of reading ±1 digit
Load section
Battery voltage
10 – 60 V DC
Max. current
270 A
Max. power
15 kW
Load patterns
Constant current, constant power, con-
stant resistance, current or power profile
Current setting
0-270.0 A (2999.9 A)
1)
Power setting
0-15.00 kW (299.99 kW)
1)
Resistance setting
0.1-2999.8 Ω
Battery voltage range 2 ranges, selected automatically at start
of test
Stabilization (For
internal current meas-
urement)
±(0.5% of reading + 0.5 A)
Battery
voltage
Highest permissible
current
Resistor ele-
ment (Nominal
values)
Range 1 10 – 27.6 V 270 A
0.069 Ω
Range 2 10 – 55.2 V 270 A
0.138 Ω
1) Maximum value for a system with more than one load unit
Inputs, maximal values
EXTERNAL
CURRENT
MEASUREMENT
EXTERNAL
CURRENT
1 V DC, 300 V DC to ground. Current
shunt should be connected to the nega-
tive side of the battery
START / STOP
Closing / opening contact
Closing and then opening the contact will
start / stop Torkel. It is not possible to keep
the contacts in closed position.
Delay until start
200 – 300 ms
Stop delay
100 – 200 ms
Battery
60 V DC, 500 V DC to ground
VOLTAGE SENSE
60 V DC, 500 V DC to ground
SERIAL
< 15 V
ALARM
250 V DC 0.28 A
28 V DC 8 A
250 V AC 8 A
Outputs, maximal values
START / STOP
5 V, 6 mA
TXL
Relay contact
SERIAL
< 15 V
ALARM
Relay contact