Atec Megger-torkel-840-860 User Manual
Page 2
![background image](/manuals/319065/2/background.png)
Programma
TORKEL 840/860 Battery Load Units
FEATURES AND BENEFITS
■
Batteries can be tested “in service”. TORKEL will adjust
to include the load current in the test parameters.
■
User adjustable alarm and shutdown points to avoid
excessive discharge.
■
No loss of test data in the event of a power loss during
testing.
■
Easily expandable for larger battery banks using TXL
extra load units.
■
Can be used “stand alone” or with a PC running
TORKEL WIN Software.
SPECIFICATIONS
Specifications are valid at nominal input voltage and an ambient
temperature of +25° C, (77° F). Specifications are subject to
change without notice.
Environment
Application field
The instrument is intended for use in high-voltage substations and
industrial environments.
Temperature
Operating
0° C to +40° C (32° F to +104° F)
Storage & transport
-40° C to +70° C (-40° F to +158° F)
Humidity
5% – 95% RH, non-condensing
CE-marking
Safety standards
IEC 61010-1:2001 Incl. national dev. for US
and CAEN 61010-1:2001
EMC standards
EN 61326: 1997+A1:1998+A2:2001
General
Power supply voltage
100 – 240 V AC, 50/60 Hz
Power
consumption (max)
150 W
Protection
Thermal cut-outs, automatic overload
protection
Dimensions
Instrument
210 x 353 x 700 mm (8.3” x 13.9” x 27.6”)
Transport case
265 x 460 x 750 mm (10.4” x 18.1” x 29.5”)
Weight
21.5 kg (47.4 lbs)38 kg (83.8 lbs) with
accessories and transport case.
Display
LCD
Available languages
English, French, German, Spanish, Swedish
Measurement section
Current measurement
Display range
0.0 – 2999 A
Accuracy
±(0.5% of reading +0.2 A)
Resolution
0.1 A
Internal current measurement
Range
0 – 270 A
Input for clamp-on ammeter
Range
0 – 1 V
mV/A-ratio
Software settable, 0.3 to 19.9 mV/A
Input impedance
>1 M
Ω
Voltage measurement
Display range 0.0 – 60 V
Accuracy
±(0.5% of reading +0.1 V)
Resolution
0.1 V
Display range 0.0 – 500 V
Accuracy
±(0.5% of reading +1 V)
Resolution
0.1 V
Time measurement
Accuracy
±0.1% of reading ±1 digit
Load section
Max. battery voltage
288 V DC (TORKEL 840)
480 V DC (TORKEL 860)
Max. current
110 A
Max. power
15 kW
Load patterns
Constant current, constant power, constant
resistance, current or power profile
Current setting
0-110.0 A (2999.9 A)
1
Power setting
0-15.00 kW (299.99 kW)
1
Resistance setting
0.1-2999.8
Ω
Battery voltage range, 4 ranges, selected automatically at
TORKEL 840
start of test
Battery voltage range, 5 ranges, selected automatically at
TORKEL 860
start of test
Stabilization
±(0.5% of reading +0.5 A)
(For internal
current measurement)
Battery
Highest
Resistor
voltage
permissible
element
current
(Nominal values)
Range 1
10 – 27.6 V
110 A
0.165
Ω
Range 2
10 – 55.2 V
110 A
0.275
Ω
Range 3
10 – 144 V
110 A
0.55
Ω
Range 4
10 – 288 V
55 A
3.3
Ω
Range 5
2
10 – 480 V
55 A (max
3.3
Ω
power 15 kW)
1) Maximum value for a system with more than one load unit
2) TORKEL 860