Specifications
Analyzers
Oscilloscope
Channels..............................................
2
Data storage, cursors, autoscaling
Max input bandwidth .....................
50 kHz
1
Max sampling rate .........................
500 kHz/channel
1
Range ..............................................
±10 V
Input resolution...............................
12, 16, or 18 bits
1
Bode Analyzer
Frequency and phase plots
Frequency range and step control
Logarithmic or linear frequency spacing
Data storage, cursors, autoscaling
Frequency range .............................
5 Hz to 35 kHz
1
Dynamic Signal Analyzer
Input range ..........................................
±10 V
Input resolution ...................................
12, 16, or 18 bits
1
Impedance Analyzer
Measurement frequency range ..........
5 Hz to 35 kHz
2-Wire Current Voltage Analyzer
Voltage range......................................
±10 V
Current range ......................................
±10 mA
3-Wire Current Voltage Analyzer
NPN BJT transistor only
Data storage, cursors, autoscaling
Maximum collector voltage............
10 V
Base current resolution ..................
1 µA (16-bit analog output)
15 µA (12-bit analog output)
Digital Multimeter
Resistance
Accuracy..............................................
1%
Range ..................................................
5
Ω to 3 M Ω
DC Voltage
Accuracy..............................................
0.3%
Range ..................................................
±20 V
Input impedance .................................
1 M
Ω
AC Voltage
Accuracy..............................................
0.3%
Range ..................................................
±14 V
rms
Current
DC accuracy ........................................
0.25% ±3 mA
2
AC accuracy ........................................
0.25% ±3 mA
2
Range ..................................................
±250 mA
Shunt resistance .................................
0.5
Ω
Maximum common-mode voltage ......
±20 V
Common-mode rejection.....................
70 dB
Capacitance
Accuracy..............................................
2%
Range ..................................................
50 pF to 500 µF
Test voltage range ..............................
1 V
pp
Continuity
Resistance threshold ..........................
15
Ω max
Inductance
Accuracy..............................................
1%
Range ..................................................
100 µH to 100 mH
Test frequency.....................................
950 Hz
Test frequency voltage........................
1 V
pp
Digital I/O
Digital input resolution .......................
8 bits
Digital output resolution.....................
8 bits
Digital addressing ...............................
4 bits
Source
Function Generator
Manual or software control
Sine, triangle, square waveforms
Frequency sweep
TTL sync pulse out
AM, FM modulation
Frequency range .............................
5 Hz to 250 kHz
Frequency accuracy ........................
3%
Output amplitude............................
±2.5 V
Software amplitude resolution ......
8 bits
Offset range....................................
±5 V
AM voltage .....................................
10 V max
Amplitude modulation ....................
Up to 100%
FM voltage......................................
10 V max
Amplitude Flatness
To 50 kHz .............................................
0.5 dB
To 250 kHz ...........................................
3 dB
Arbitrary Waveform Generator
Channels..............................................
2, 1-shot or
continuous generation
1
Specification depends on data acquisition device functionality.
2
Proper null correction at the common-mode voltage can reduce ±3 mA error to 200 µA noise.
NI Educational Laboratory Virtual Instrumentation Suite (NI ELVIS)
2