Measurement accuracy – Atec Fluke-2680 Series User Manual
Page 6
6 Fluke Corporation Fluke 2680 Series Data Acquisition Systems
1
Total instrument accuracy for the indicated time period and ambient temperature
range. Includes A/D errors, linearization conformity, initial calibration error,
isothermality errors, reference junction conformity and power line voltage effects
within the range from 100VAC to 264VAC.
2
Sine wave inputs >2000 counts (slow), >200 counts (fast). Accuracies for crest
factor ≤2.0.
3
For two-wire measurements add 5Ω to basic accuracy (does not include lead-
wire resistances).
4
For two-wire measurements add 700-1000Ω to basic accuracy (does not include
lead-wire resistances). Ohms varies due to the resistance of the solid state
switches.
5
DIN/IEC 751 only, assumes no lead-wire resistance errors.
6
Resolution is 0.02°C or 0.04°F over the useful range of base metal thermocouples
(J, K, T, E, N, L, U) and 0.1°C or 0.2°F resolution for types R, S, B, and C with
slow scan.
7
Resolution is 0.2°C or 0.4°F over the useful range of base metal thermocouples (J,
K, T, E, N, L, U) and 1.0°C or 2.0°F resolution for types R, S, B, and C with slow
scan.
8
Accuracy for both slow and fast scan speeds.
9
Open thermocouple detection is performed on each thermocouple channel unless
defeated by computer command.
10
Using Stein hart – Hart thermistor polynomial: T = A+B (InR) + C (InR)
3
T = temp in °K
A, B and C = fitting constants
R = resistance of thermistor in Ω.
Measurement accuracy
cont.
Model 2680A-PAI
AC Voltage
Range
Resolution
Frequency
Accuracy
1,2
,3σ
σ
σ
σ
σ,± (% input+counts)
Slow
Fast
300 mV
1 µV
20 Hz-50 Hz
3.0%+25
6.0%+50
50 Hz-20 kHz
0.4%+25
1.0%+50
20 kHz-50 kHz
2.0%+30
3.0%+50
50 kHz-100 kHz
5.0%+50
5.0%+100
3V
100 µV
Same frequencies, similar accuracies as above
30V
1 mV
Same frequencies, similar accuracies as above
150/300V 10 mV
Same frequencies, similar accuracies as above
RTD (Pt 100)
Accuracy
1,5
, 3σ
σ
σ
σ
σ, ± °C (4-wire)
Temperature
Resolution
90 Day,
1 Year,
°C
°C
18 to 28°C
18 to 28°C
Slow
Slow
Slow
-200
0.003
0.06
0.09
0
0.003
0.09
0.13
100
0.003
0.10
0.16
300
0.003
0.14
0.21
600
0.003
0.19
0.30
Thermistor
10
2 k to 100 kΩ
-40 °C to 150 °C
0.003
0.3
0.4
Frequency Measurement Accuracy
1,8
, -20 to 60°C
Range
Resolution
Accuracy, 3σ
σ
σ
σ
σ,± (% input +Hz)
Slow
Fast
Slow
Fast
15 Hz-900 Hz
0.01 Hz
0.1 Hz
0.05%+0.02 Hz
0.05%+0.2 Hz
900 Hz-9 kHz
0.1 Hz
1 Hz
0.05%+0.1 Hz
0.05%+1 Hz
9 kHz-90 kHz
1 Hz
10 Hz
0.05%+1 Hz
0.05%+10 Hz
90 kHz-900 kHz
10 Hz
100 Hz
0.05%+10 Hz
0.05%+100 Hz
1 MHz
100 Hz
1 kHz
0.05%+100 Hz
0.05%+1 kHz
Frequency Measurement Sensitivity (sine wave)
Frequency
Minimum Signal
Maximum Signal
15 Hz - 200 Hz
100 mV rms
150/300V rms
200 Hz - 70 kHz
100 mV rms
30V rms
70 kHz - 100 kHz
100 mV rms
20V rms
100 kHz - 200 kHz
150 mV rms
10V rms
200 kHz - 300 kHz
150 mV rms
7V rms
300 kHz - 1 MHz
linearly increasing from
linearly decreasing from
150 mV rms at 300 kHz to 7V rms at 300 kHz to
2V rms at 1 MHz
2V rms at 1 MHz
Model 2680A-FAI
AC Voltage
Range
Resolution
Frequency
Accuracy
1,2
,3σ
σ
σ
σ
σ,± (% input+counts)
Slow
Fast
300mV
10 µV
20 Hz-50 Hz
3.0%+25
6.0%+50
50 Hz-20 kHz
0.4%+25
1.0%+50
20 kHz-50 kHz
2.0%+30
3.0%+50
50 kHz-100 kHz
5.0%+50
5.0%+100
3V
100 µV
Same frequencies, similar accuracies as above
30V
1 mV
Same frequencies, similar accuracies as above
RTD (Pt 100)
Accuracy
1,5
, 3σ
σ
σ
σ
σ, ± °C (4-wire)
Temperature
Resolution
90 Day,
1 Year,
°C
°C
18 to 28°C
18 to 28°C
Slow
Slow
Slow
-200
0.03
0.16
0.25
0
0.03
0.20
0.31
100
0.03
0.23
0.34
300
0.03
0.30
0.41
600
0.03
0.53
0.63
Thermistor 10 k to 100 kΩ
-40 °C to 150 °C
0.03
0.4
0.5
Frequency Measurement Accuracy
1,8
, -20 to 60°C
Range
Resolution
Accuracy, 3σ
σ
σ
σ
σ,± (% input +Hz)
Slow
Fast
Slow
Fast
15 Hz-900 Hz
0.01 Hz
0.1 Hz
0.05%+0.02 Hz
0.05%+0.2 Hz
900 Hz-9 kHz
0.1 Hz
1 Hz
0.05%+0.1 Hz
0.05%+1 Hz
9 kHz-90 kHz
1 Hz
10 Hz
0.05%+1 Hz
0.05%+10 Hz
90 kHz-900 kHz
10 Hz
100 Hz
0.05%+10 Hz
0.05%+100 Hz
1 MHz
100 Hz
1 kHz
0.05%+100 Hz
0.05%+1 kHz
Frequency Measurement Sensitivity (sine wave)
Frequency
Minimum Signal
Maximum Signal
15 Hz - 200 Hz
100 mV rms
30V rms
200 Hz - 70 kHz
100 mV rms
30V rms
70 kHz - 100 kHz
100 mV rms
20V rms
100 kHz - 200 kHz
150 mV rms
10V rms
200 kHz - 300 kHz
150 mV rms
7V rms
300 kHz - 1 MHz
linearly increasing from
linearly decreasing from
150 mV rms at 300 kHz to 7V rms at 300 kHz to
2V rms at 1 MHz
2V rms at 1 M