beautypg.com

High power system sourcemeter smu instrument, 2657a – Atec Keithley_2657A User Manual

Page 4

background image

S

M

U

I

N

S

TR

U

M

E

N

TS

A Greater Measure of Confidence

www.keithley.com

1.888.KEITHLEY

(U.S. only)

2657a

High Power System SourceMeter

SMU Instrument

TrIGGErING aND sYNCHroNIZaTIoN sPECIfICaTIoNs

TRIGGERING: Trigger In to Trigger Out: 0 .5µs, typical .
SyNCHRONIzATION: Single- or multi-node synchronized source change: <0 .5µs, typical .

ProGraMMING

TEST SCRIPT BUILDER: Integrated development environment for building, running, and

managing TSP scripts .

TSP EXPRESS (Embedded): Tool that allows users to perform common I-V tests quickly and easily

without programming or installing software .

SOFTWARE INTERFACE: TSP Express (Embedded), Direct GPIB/VISA, Read/Write with VB,

VC/C++, VC#, LabVIEW™, TestPoint™, LabWindows™/CVI, etc .

sYsTEM EXPaNsIoN

The TSP-Link expansion interface allows TSP-enabled instruments to trigger and communicate
with each other . See figure below:

GENEral

USB: USB 2 .1 Host Controller, supports external data storage .
CONTACT CHECK: ±50

W .

PC INTERFACE: IEEE-488 .1 and .2; LXI Ethernet; RS-232 .
DIGITAL I/O INTERFACE: Input/Output Pins: 14 open drain I/O bits . 5 .25V max .
POWER SUPPLy: 100V to 250VAC, 50Hz–60Hz (auto sensing), 550VA max .
COOLING: Forced air . Side and top intake and rear exhaust .
EMC: Conforms to European Union EMC Directive .
SAFETy: ETL listed (PENDING) . Conforms to European Union Low Voltage Directive .
WARRANTy: 1 year .
DIMENSIONS: 89mm high × 435mm wide × 549mm deep (3 .5 in × 17 .1 in × 21 .6 in) . Bench

Configuration (with handle and feet): 104mm high × 483mm wide × 620mm deep (4 .1 in
× 19 in × 24 .4 in) .

WEIGHT: 9 .98kg (22 lbs) .
ENVIRONMENT: For indoor use only .
CALIBRATION PERIOD: One year .

To Nodes 3–32

Node 1

Node 2

To Host Computer

M

ode

l 2

65

7A s

pe

ci

fic

at

io

ns

M

ode

l 2

65

7A s

pe

ci

fic

at

io

ns