capabilities, full level control and
measurement facilities of modulation, and
precision power meter features for
enhanced sensitivity and high accuracy
testing needs.
Analog paging encoding/decoding,
DTMF, tone coded squelch, digital squelch,
AM modulation/demodulation with two
separate AF generators and a cross band
duplex feature gives you added test
versatility in a variety of public and private
wireless systems.
Software Options Make Complex
Testing Simple
As with every IFR test set, you get the
advantage of IFR applications engineering
support.
Our comprehensive portfolio of
application software options is designed to
automate and expand the functionality of
your instrument.
Plus, on-going software support means
that you can easily upgrade your
FM/AM-1600S/CSA when standards
definitions or test and service requirements
change.
AC1009W - EasySpan is a Windows-
based software utility which extracts
spectrum analyzer and tracking
generator traces from the IFR
FM/AM-1600S/CSA to a PC for further
off-line analysis.
AC1011 - Simultaneously analyzes the
signal strength or channel activity of up
to three voice channel groups or 100
frequencies.
AC1012 - AutoCell 882 is a
comprehensive test package for Ericsson
model 882 base station transceivers.
AC1017 - AutoCell-Series II is a
comprehensive program for FCC
compliance testing of Lucent Series II
cell sites.
AC1018 - Provides automated testing
and calibration of Motorola HD-II/LD cell
sites.
AC1019 - EasySweep is a swept
measurement utility designed to test
antennas and transmission lines.
AC1020D - AutoCell NT and AutoCell
NTD provides automated testing of
Northern Telecom cell sites.
Dynamic IQ Constellation Display
Simplifies Analysis
The FM/AM-1600S/CSA provides you
with a dynamic, high speed constellation
display for precise RF modulation analysis
of
π
/4 DQPSK digitally modulated
waveforms from 10 MHz to 1 GHz. This
unique FM/AM-1600S/CSA feature gives a
near-real time display of digital modulation
characteristics that points out the cause of
the trouble in digital radios.
IQ Constellation display allows for comprehensive
digital modulation testing
Complex Functionality Made Easy
Even with its elaborate capabilities, the
FM/AM-1600S/CSA was developed to
execute complex tests simply and with
minimal operator training.
Using field-proven front panel and
user man-machine interfaces, the
FM/AM-1600S/CSA gives you the
performance and ease-of-use features that
significantly reduce your testing and
training time.
Its test macro command programming
language (TMAC) can be easily configured
to perform complex automatic tests on
base stations and terminals (including
single and multi-mode as well as single
and multi-band radios), and baseband
equipment. This powerful capability allows
you to create and save simple "one button"
test routines for future use. This flexibility
means you can create and execute
complex and repeatable routines no matter
what your level of expertise.
A color display gives FM/AM-1600S/CSA
users vivid screen clarity. Extensive use of
softkeys reduces your complex cellular and
PCS parametric and protocol tests to fast,
simple and manageable routines.
Specification
(T/R & DUPLEX CONNECTORS)
Frequency Range
400 kHz to 999.9999 MHz
Resolution
100 Hz
Accuracy
Same as Master Oscillator
Output Level Range (T/R)
-137 dBm to 0 dBm into 50
Ω
(-40 dBm maximum with reverse power present)
Duplex Level Range
-120 dBm to + 7 dBm or into 50
Ω
Duplex Resolution
0.1 dB
Level Accuracy (T/R)
±1.5 dB (level
≥
-110 dBm)
±2.5 dB (levels from <-110 dBm to -127 dBm)
Duplex Level Accuracy
±1.5 dB (level
≥
-100 dBm)
±2.5 dB (levels from <-100 dBm to -120 dBm)
VSWR (T/R)
<1.35:1 typical
Duplex VSWR
<1.38:1
INPUT PROTECTION
(T/R)
See RF Power Meter
(DUPLEX)
65 W (15 seconds with alarm)
SPECTRAL PURITY
Harmonics
<-26 dBc
Non-Harmonics
<-50 dBc
Residual FM
<10 Hz (RMS, 0.3 kHz to 3 kHz BW)
Residual AM
<0.05% (RMS, 0.3 kHz to 3 kHz BW)
SSB Phase Noise
<-90 dBc/Hz (25 kHz offset)
<-85 dBc/Hz (25 kHz offset, f
c
≥
930 MHz)
<-80 dBc/Hz (25 kHz offset, f
c
<1 MHz)
Internal FM Range
Off, 100 Hz to 100 kHz Dev
Accuracy
±5% (1 kHz to 25 kHz Dev, 1 kHz rate)
±7% (<1 kHz Dev and >25 kHz Dev, 1 kHz rate)
Resolution
100 Hz (
≤
25 kHz Dev)
500 Hz (>25 kHz Dev)
Rate
Off, 30 Hz to 40 kHz
(
≤
20 kbs digital)
Waveforms
Sine, Square, Triangle, Ramp and Pulse
Distortion
<1 % (1 kHz rate, 0.3 kHz to 3 kHz BW,
1 kHz to 25 kHz deviation)
INTERNAL AM RF GENERATOR
Range
500 kHz to 999.9999 MHz
Internal AM Range
Off, 1% to 90%
Accuracy
±5% of setting (30% to 90% modulation, 1 kHz
sinewave)
±10% of setting (30% to 90% modulation, 1 kHz
sinewave f
c
< 1 MHz)
Resolution
1 %
Rate
100 Hz to 10 kHz
Waveforms
Sine, Square, Triangle, Ramp and Pulse
Distortion
<1% (30% to 70% modulation, 1 kHz sinewave, f
c
>1 MHz, 0.3 kHz to 3 kHz BW)
Internal Phase Modulation Range
Off, 0.1 to 10 Radians
Accuracy
±10% (at 1 kHz rate, > 0.3 Rad)
Resolution
0.1 Radians
Rate
Off, 30 Hz to 6 kHz
Waveforms
Sine, Square, Triangle, Ramp and Pulse
Distortion
<2% (1 kHz rate, 0.3 kHz to 3 kHz BW,
≥
0.5 Rad)
External
External inputs with the same characteristics as the
internal modulation sources are supported.
A 10 Vp-p injection level is required to obtain
indicated setting ±10%.
Modulation
RF Signal Generator
FM/AM-1600S/CSA
FM/AM-1600S/CSA
2
http://www.ifrinternational.com