Specifications

1

(Preliminary)

External Power Meter

FOT-932

FOT-932X

FOT-933

Detector type

Ge

GeX

InGaAs

Measurement range (dBm)

10 to -70

26 to -55

6 to –73

Range displayed (dBm)

Down to -90

Down to -75

Down to -90

Uncertainty

2, 3

± 5 % (± 0.1 nW)

± 5 % (± 3 nW)

± 5 % (± 0.05 nW)

Wavelength range (nm)

800 to 1650

800 to 1650

800 to 1650

Display resolution

2

(dB)

0.01

0.01

0.01

Calibrated wavelengths

40

40

40

Recommended recalibration
period (years)

3

3

3

Automatic offset nulling

4

Yes

Yes

Yes

Sources

Standard

-4

-5

-12C (second port)

-12D (second port)

Wavelengths (nm)

1310 ± 20

1310 ± 20

1310 ± 20

850 ± 25

850 ± 25

1550 ± 20

1550 ± 20

1490 ± 10

1300 +50/–10

1300 +50/–10

1625 ± 15

1550 ± 20

Emitter type

Laser

Laser

Laser

LED

LED

Minimum output power

5

(dBm)

-3.5/-3.5

-3.5/-7/-7

-3.5/-7/-7

-26.5/-26.5 (50/125 µm)

-24.5/-24.5 (62.5/125 µm)

Spectral width

6

(nm)

5/

5

5/

5/

5

5/

5/

5

50/135

50/135

Stability

7

(8 hours) (dB)

± 0.05

± 0.05

± 0.05

± 0.05

± 0.05

FasTest

Standard

-4

-5

-12C (second port)

-12D (second port)

Wavelengths (nm)

1310 ± 20

1310 ± 20

1310 ± 20

850 ± 25

850 ± 25

1550 ± 20

1550 ± 20

1490 ± 10

1300 +50/–10

1300 +50/–10

1625 ± 15

1550 ± 20

Loss range

8

(dB)

60

60

60

43

45

Loss precision

9

(repeatability) (dB)

Side-by-side

0.15

0.15

0.15

0.25

0.25

Loopback

0.25

0.25

0.25

0.45

0.45

Length measurement range (km)

200

200

200

5

5

Length measurement uncertainty

10

± (10 m + 1 % x distance)

Dedicated ORL

All SM wavelengths

ORL range

9

(APC / UPC) (dB)

65 / 55

ORL uncertainty

11

(dB)

± 0.5

Resolution

2

(dB)

0.01

General Specifications

Size (H x W x D)

25.0 cm x 12.5 cm x 7.5 cm

(9

7

/

8

in x 4

15

/

16

in x 3 in)

Weight 1

kg

(2.2

lb)

Temperature

operating

–10 °C to 50 °C

(14 °F to 122 °F)

storage

–40 °C to 70 °C

(–40 °F to 158 °F)

Storage

capacity 1024 complete tests

Relative humidity

0 % to 95 % non-condensing

Power

9

Li-ion battery 9 hours

3 hours to fully recharge when unit is off

Warranty (years)

3

FOT-930 MaxTester

Notes:
1. At 23 °C ± 1 °C and 1550 nm with FC/PC connector unless otherwise specified.
2. Resolution, uncertainty and linearity are functions of input power; uncertainty is

valid at calibration conditions.

3. Traceable to NIST; up to 20 dBm for GeX.
4. Power of > –45 dBm for Ge, > –30 dBm for GeX and > –57 dBm for InGaAs.
5. In high-power mode.
6. As defined by Telcordia TR-TSY-000887, rms for lasers and at -3 dB for LEDs;

typical values for LEDs.

7. After a warm-up time of 6 minutes.
8. Typical value, at 1550 nm for SM and 850 nm for MM.
9. Typical value.
10. For fiber length

120 km.

11. Typical value, for APC connectors.
12. In graded-index MM fibers.

Talk set

Emitter type

Laser

Wavelength (nm)

1550 ± 20

Dynamic range at 1550 nm (dB)

45

Dynamic range MM

12

(dB)

40

VFL

9

Emitter type

Laser

Wavelength (nm)

650

Output power (dBm)

3

Standard Accessories

User guide, AC adapter/charger, 2 Li-ion batteries, shoulder strap, Certificate of Calibration, connector adapter (FOA) according to chosen connector, mandrel,
RS-232 cable and software, soft carrying case and alcohol cleaning pads.