Version 06.00, April 2009

4

Rohde & Schwarz R&S

®

CBT/R&S

®

CBT32 Bluetooth

®

Tester

Output level uncertainty

in temperature range +20 °C to +35 °C

output level < –10 dBm

< 1.0 dB

output level

≥ –10 dBm

< 1.5 dB

in temperature range +5 °C to +45 °C

output level < –10 dBm

< 1.5 dB

RF IN/OUT

output level

≥ –10 dBm

< 2.0 dB

Output level resolution

0.1

dB

Generator RF output level repeatability typical values after 1 h warm-up time at

constant ambient temperature

< 0.03 dB

VSWR
RF IN/OUT

< 1.5

Attenuation of harmonics

f

0

= 2398 MHz to 2499 MHz, up to 7 GHz

RF IN/OUT

> 30 dB

Attenuation of nonharmonics

> 50 dB

GFSK modulation
GFSK bit rate

DHx packet types

1 Mbps, B × T = 0.5

Modulation index

11110000 pattern
frequency deviation 160 kHz

0.32

Modulation index range

frequency deviation 100 kHz to 220 kHz

0.20 to 0.44

Modulation index resolution

0.01

Modulation index uncertainty

11110000 pattern
frequency deviation 160 kHz
in temperature range +20 °C to +35 °C

±1 %

DPSK modulation
π/4DQPSK bit rate

2-DHx packet types

2 Mbps

8DPSK bit rate

3-DHx packet types

3 Mbps

Symbol rate

1 Msps

Modulation uncertainty

DEVM

≤ 5 %, rms

GFSK modulation Bluetooth

®

Low Energy

GFSK bit rate

RF_PHY_Test_Ref, ADV_IND,
ADV_NONCONN_IND and DATA packet
types

1 Mbps, B × T = 0.5

Modulation index

11110000 pattern
frequency deviation 250 kHz

0.50

Modulation index range

frequency deviation 200 kHz to 300 kHz

0.40 to 0.60

Modulation index resolution

0.01

Modulation index uncertainty

11110000 pattern
frequency deviation 250 kHz
in temperature range +20 °C to +35 °C

±1 %

Dirty TX (basic rate)
Frequency offset range

±250 kHz

Frequency offset resolution

1 kHz

Frequency offset uncertainty

±5 Hz + drift of time base

Modulation index range

0.20 to 0.44

Modulation index resolution

0.01

Modulation index uncertainty

11110000 pattern
in temperature range +20 °C to +35 °C

±1 %

Symbol time error range

±20 ppm

Symbol time error resolution

1 ppm