beautypg.com

Electric and magnetic field measurement, Specifi cations – Atec Narda-EHP-50C User Manual

Page 4

background image

35

narda

Safety Test Solutions an

communications company USA • Germany • Italy

USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: [email protected] • www.narda-sts.us

Electric and Magnetic Field Measurement

EHP-50C Electric Field and Magnetic Flux Density Measurement

Specifi cations

EHP-50C E and

β Field Analyzer

FUNCTIONAL SPECIFICATIONS

Electric Field

Magnetic Field

Frequency range

5 Hz to 100 kHz

Measuring Ranges

1 kV/m to 100 kV/m

0.1 mT to 10 mT

Overload

200 kV/m @ 60 Hz

20 mT @ 60 Hz

Resolution

0.1 V/m

10 nT

Sensitivity

0.01 V/m

1 nT

Absolute Error

±0.5 dB @ 60 Hz and 1 kV/m

±0.5 dB @ 60 Hz and 0.1 mT

Flatness (40 Hz – 10 kHz)

±0.5 dB

±0.5 dB

Isotropy

±1 dB

Linearity @ 60 Hz

±0.2 dB (1 V/m to 100 kV/m)

±0.2 dB (200 nT to 10 mT)

SPAN

100 Hz, 200 Hz, 500 Hz, 1 kHz, 2 kHz, 10 kHz, 100 kHz

Starting Frequency

1.2 % of the SPAN

Stop Frequency

Equal to the SPAN

E-Field Rejection

> 20 dB

H-Field Rejection

> 20 dB

FFT

Real Time FFT Analysis

Internal Data Logger

1 measurement every 30 or 60 seconds

Internal memory

1440 data with 1 minute storing, 2880 data with 30 second storing.

The data can only be transferred to a PC

GENERAL SPECIFICATIONS

Calibration

Internal EEPROM

Temperature Error (referred to 23°C)

±0.05 dB/°C between -10 and +23°C @ 40% RH

±0.01 dB/°C between +23 and +50°C @ 40% RH

Humidity Error (referred to 40%)

±0.05 dB between 20% and 50% @ +23°C

±0.05 dB between 50% and 80% @ +23°C

Internal Battery

Rechargeable NiMH batteries (5 x 1.2 V)

Operating Time

> 10 hours in normal mode

> 150 hours in low power mode

24 hours with internal data logger (SPAN > 200 Hz) in stand alone mode

Recharging Time

< 4 hours

External DC supply

10 / 15 V, 200 mA

Optical Fiber Link and operating Distance

Up to 80 meters

Firmware Update

Via USB port

Operational and Storage Temperature

-10°C / 50°C – Storage -20°C / 70°C

Size and Weight

92 x 92 x 109 mm – 25 g

Tripod Support

Threaded insert π

ORDERING INFORMATION

Part Number

EHP-50C

Accessories Supplied

0.5m Tripod, Optical Shorting Loop, EHP Software,

Operating Manual and Calibration Certificate