Electric and Magnetic Field Measurement
EHP-50C Electric Field and Magnetic Flux Density Measurement
Specifi cations
EHP-50C E and
β Field Analyzer
FUNCTIONAL SPECIFICATIONS
Electric Field
Magnetic Field
Frequency range
5 Hz to 100 kHz
Measuring Ranges
1 kV/m to 100 kV/m
0.1 mT to 10 mT
Overload
200 kV/m @ 60 Hz
20 mT @ 60 Hz
Resolution
0.1 V/m
10 nT
Sensitivity
0.01 V/m
1 nT
Absolute Error
±0.5 dB @ 60 Hz and 1 kV/m
±0.5 dB @ 60 Hz and 0.1 mT
Flatness (40 Hz – 10 kHz)
±0.5 dB
±0.5 dB
Isotropy
±1 dB
Linearity @ 60 Hz
±0.2 dB (1 V/m to 100 kV/m)
±0.2 dB (200 nT to 10 mT)
SPAN
100 Hz, 200 Hz, 500 Hz, 1 kHz, 2 kHz, 10 kHz, 100 kHz
Starting Frequency
1.2 % of the SPAN
Stop Frequency
Equal to the SPAN
E-Field Rejection
—
> 20 dB
H-Field Rejection
> 20 dB
—
FFT
Real Time FFT Analysis
Internal Data Logger
1 measurement every 30 or 60 seconds
Internal memory
1440 data with 1 minute storing, 2880 data with 30 second storing.
The data can only be transferred to a PC
GENERAL SPECIFICATIONS
Calibration
Internal EEPROM
Temperature Error (referred to 23°C)
±0.05 dB/°C between -10 and +23°C @ 40% RH
±0.01 dB/°C between +23 and +50°C @ 40% RH
Humidity Error (referred to 40%)
±0.05 dB between 20% and 50% @ +23°C
±0.05 dB between 50% and 80% @ +23°C
Internal Battery
Rechargeable NiMH batteries (5 x 1.2 V)
Operating Time
> 10 hours in normal mode
> 150 hours in low power mode
24 hours with internal data logger (SPAN > 200 Hz) in stand alone mode
Recharging Time
< 4 hours
External DC supply
10 / 15 V, 200 mA
Optical Fiber Link and operating Distance
Up to 80 meters
Firmware Update
Via USB port
Operational and Storage Temperature
-10°C / 50°C – Storage -20°C / 70°C
Size and Weight
92 x 92 x 109 mm – 25 g
Tripod Support
Threaded insert π
ORDERING INFORMATION
Part Number
EHP-50C
Accessories Supplied
0.5m Tripod, Optical Shorting Loop, EHP Software,
Operating Manual and Calibration Certificate