beautypg.com

Technical specifications – Atec Narda-EHP-50D User Manual

Page 8

background image

EHP-50D 04-11

8 / 12

Subject to change without notice



TECHNICAL SPECIFICATIONS

EHP-50D

FUNCTIONAL SPECIFICATIONS

(1)

Electric Field

Magnetic Field

Frequency Range

5 Hz ÷ 100 kHz

Measurement Ranges

(2)

5 mV/m ÷ 1 kV/m

500 mV/m ÷ 100 kV/m

(total measurement range: 146 dB)

0.3 nT ÷ 100 µΤ

30 nT ÷ 10 mT

(total measurement range: 150 dB)

Overload

200 kV/m

20 mT

Dynamic range

106 dB

110 dB

Resolution

(3)

1 mV/m with NBM-550

0.1 mV/m with EHP-TS software

1 mV/m in Stand Alone mode

0.1 nT with NBM-550

0.1 nT with EHP-TS software

1 nT in Stand Alone mode

Displayed Average Noise Level

(4)

Isotropic result
Single axis

5 mV/m
3 mV/m

0.3 nT
0.2 nT

Flatness (@ 100V/m and 2µΤ)
(5 Hz ÷ 40 Hz)
(40 Hz ÷ 100 kHz)

0.8 dB

0.35 dB

0.8 dB

0.35 dB

Typical 3D anisotropy

(5)

0.54 dB

0.12 dB

Linearity

(6)

0.2 dB (1 V/m ÷ 1 kV/m)

0.2 dB ( 200 nT ÷ 10 mT)

SPAN

100 Hz, 200 Hz, 500 Hz, 1 kHz, 2 kHz, 10 kHz, 100 kHz

(500 Hz to 100 kHz in Stand Alone mode)

Start frequency

1.2 % of the SPAN

Stop frequency

Equal to the SPAN

Rejection to E fields

---

> 20 dB

Rejection to H fields

> 20 dB

---

Spectrum analysis method

FFT

Acquisition method

Simultaneous three-axis acquisition

Internal data logger

1 measurement every 30 or 60 seconds

Internal memory

(7)

Up to 24 hours regardeless of the logging rate

(1) Unless otherwise indicated, specifications are referred to an ambient operating temperature of 23°C and relative humidity of 50%
(2) For each individual axis. Ranges to be selected manually
(3) For the lower measurement range
(4) DANL is frequency and SPAN dependent. The specified best performance is referred to f ≥ 50Hz and SPAN ≤ 1kHz
(5) Typical value @ 50Hz, see application note "BG_0509_ELF_measurements_uncertainty” for details
(6) Referred to 100 V/m and 1 µT @ 50 Hz
(7) Measurement results acquired in stand alone mode

can be transferred to PC only