Personal and area monitors, Specifications for indoor models, Alarm operation control input – Atec Narda-8800 Series User Manual
Page 3: T.o.c

Personal and Area Monitors
///////////
435 MORELAND ROAD HAUPPAUGE, NY 11788
USA TEL: 516 231-1700 INTL TEL: (1) 516 231-1390 FAX (1) 516 231-1711
E-MAIL: [email protected] www.nardaeast.l-3com.com
SPECIFICATIONS FOR INDOOR MODELS
MODEL No.
8810
8815
8820
8825B
Frequency Range
2 to 30 MHz
10 to 500 MHz
0.5 to 18 GHz
0.5 to 100 GHz
Frequency Sensitivity
±1.5 dB
±
2.0 dB
3.0 dB
=
4.5 dB
=
0.5 - 50 GHz
6.0 dB
=
0.5 - 100 GHz
Alarm Threshold
10 % of Standard
>
0.1 mW/cm
1.0 mW/cm
0.5 mW/cm
Overload
CW
Peak
60 mW/cm
60 W/cm
300 mW/cm
100 W/cm
3 dB Intercept Angle
>180
°
>180
°
>180
°
?
>180
°
?
Temperature
Operating
Non-operating
-10
°
C to +55
°
C
-40
°
C to + 75
°
C
Batteries
Circuit
@
Fail-safe
A
9V lithium (supplied)
3V lithium (supplied)
Weight (approx)
24 oz. (0.7 kg)
Calibration Frequencies
B
2, 3, 10 and 30 MHz
10, 30, 100, 300,
400 and 500 MHz
8.2
B
GHz
8.2
B
GHz
Accessories Supplied
Operation Manual, J1 and J2 Mating Connectors
Notes:
=
Peak to peak. The threshold can be factory set to provide optimum performance over a specific band of interest.
>
The alarm threshold is proportionate to IEEE C95.1-1991 Standard and varies over the 2 to 30 MHz band. The threshold is 0.9% of
the H field Maximum Permissable Exposure (MPE) from 3 to 30 MHz and equal to 10% of the E field MPE from 2 to 30 Mhz.
(Controlled Environments)
?
The 3 dB intercept angle is >180
°
with the E field tangential to the mounting plane and >165
°
with the E field perpendicular to the
mounting plane.
@
The rated life is two years. It is recommended that this battery be replaced annually during calibration.
A
The fail-safe battery has a rated life of 5 to 10 years. It is recommended that it be checked annually during calibration and replaced
every three years.
B
Other calibration frequencies available for narrow band applications. Contact the factory.
ALARM OPERATION
CONTROL INPUT
CONDITION
AUDIBLE &
VISUAL INDICATIONS
J1
OUTPUT
Normal Operation
Light flashes every 40 sec.
Low
Low Battery
Alarm chirps, light flashes
Low
TestButtonDepressed
Steady sound and light
High
Alarm
Steady sound and light
High
Main Battery Failure
No sound or light
High
Sensor Failure
Steady sound and light
High
CONDITION
J2 INPUT
Normal Operation
Open Circuit
Alarm Disabled
Short Circuit
People Present
!
Open Circuit
No People Present
!
Short Circuit
Notes:
The STATUS signal available at J1 is a voltage source that
provides >2.4 Vdc into a 500 k
Ω
load when High.
The voltage drops to <0.8 Vdc when Low. The load
impedance can be reduced to 24 k
Ω
providing that
the Main Battery Failure signal is not required.
Jack type is 3.5mm miniature phone.
J2 Input: Normal condition is an open circuit.
A short circuit (<100 ohms) disables the SMARTS.
Jack type is 2.5 mm miniature phone.
!
When used with the Model 8808 Personnel Sensor.
$&