beautypg.com

Specifications, Frequency specifications, Amplitude specifications – Atec Agilent-E4411B-03B-08B User Manual

Page 10

background image

10

Specifications

Frequency specifications

Frequency range
E4411B

50

9 kHz to 1.5 GHz

75

Ω(Opt. 1DP)

1 MHz to 1.5 GHz

E4403B

9 kHz to 3.0 GHz

E4408B

9 kHz to 26.5 GHz

Band

LO harmonic = N

0

1

9 kHz to 3.0 GHz

1

1

2.85 GHz to 6.7 GHz

2

2

6.2 GHz to 13.2 GHz

3

4

12.8 GHz to 19.2 GHz

4

4

18.7 GHz to 26.5 GHz

Frequency reference
Aging rate

±2x10

-6

/year, ±1.0x10

-7

/day,

characteristic

Settability

±5x10

-7

Temperature stability

±5x10

-6

Frequency readout accuracy
(Start, Stop, Center, Marker)

±(frequency readout x frequency
reference error

1

+ 0.75% of span

+ 15% of RBW + 10 Hz + 1Hz x N

2

)

Marker frequency counter
Accuracy

±(marker frequency x frequency
reference error

1

+ counter resolution)

Resolution

Selectable from 1 Hz to 100 kHz

Frequency span
Range

0 Hz (zero span), and

E4411B

100 Hz to 1.5 GHz

E4403B

100 Hz to 3.0 GHz

E4408B

100 Hz to 26.5 GHz

Resolution

2 Hz x N

2

Accuracy

±1% of span

Sweep time
Range

4 ms to 4000 sec.

Accuracy

±1%

Sweep trigger

Free Run, single, line, video, offset,
delayed trigger, and external

Offset trigger range

± 327 ms to ± 323 Ks

Sweep (trace) points

401

Resolution bandwidth
Range

(–3 dB bandwidth)

1 kHz to 3 MHz in 1-3-10 sequence
and 5 MHz

(–6 dB bandwidth)

9 kHz and 120 kHz

Accuracy

1 kHz to 3 MHz RBW

±15%

5 MHz RBW

±30%

Selectivity

60 dB/3 dB bandwidth ratio

<15:1, characteristic

Video bandwidth range
(–3 dB bandwidth)

30 Hz to 1 MHz in 1-3-10 sequence,
3 MHz, characteristic

Stability

Noise sidebands (1 kHz RBW, 30 Hz VBW and sample detector)E4411B

≥10 kHz offset from CW signal ≤–90 dBc/Hz

≥20 kHz offset from CW signal ≤–100 dBc/Hz

≥30 kHz offset from CW signal ≤–102 dBc/Hz

≥100 kHz offset from CW signal ≤–112 dBc/Hz
E4403B, E4408B

≥10 kHz offset from CW signal ≤–90 dBc/Hz + (20 Log N

2

for frequencies > 6.7 GHz)

≥20 kHz offset from CW signal ≤–98 dBc/Hz + 20 Log N

2

≥30 kHz offset from CW signal ≤–100 dBc/Hz + 20 Log N

2

≥100 kHz offset from CW signal ≤–112 dBc/Hz + 20 Log N

2

Residual FM

1 kHz RBW, 1 kHz VBW

≤150 Hz peak-to-peak x N

2

in 100 ms

System-related sidebands

≥30 kHz offset from

≤–65 dBc + (20 Log N

2

for frequencies > 6.7 GHz)

CW signal

Amplitude specifications

Absolute amplitude accuracy
Overall amplitude accuracy

3

±(0.6 dB +absolute frequency response)

20 °C to 30 °C

At reference settings

6

±0.4 dB

Measurement range

Displayed average noise level
to maximum safe input level

Input attenuator range
E4411B

0 to 60 dB, in 5 dB steps

E4403B, E4408B

0 to 65 dB, in 5 dB steps

Maximum safe input level
Average continuous power
E4411B (

≥15 dB attenuation)

+30 dBm (1W)

E4403B, E4408B

(

≥30 dB attenuation)

+30 dBm (1W)

Peak pulse power
E4411B (

≥15 dB attenuation)

+30 dBm (1W)

E4403B, E4408B

(

≥30 dB attenuation)

+50 dBm (100W)

1-dB gain compression (total power at input mixer)

4, 5

E4411B 0 dBm
E4403B 0 dBm
E4408B

50 MHz to 6.7 GHz

0 dBm

6.7 GHz to 13.2 GHz

–3 dBm

13.2 GHz to 26.5 GHz

–5 dBm

Displayed average noise level
(Input terminated, 0 dB attenuation, sample detector, reference level =
–70 dBm, 1 kHz RBW, 30 Hz VBW)
E4411B

400 kHz to 10 MHz

≤–115 dBm

10 MHz to 500 MHz

≤–119 dBm

500 MHz to 1.0 GHz

≤–117 dBm

1.0 GHz to 1.5 GHz

≤–113 dBm

E4411B (Option 1DP)

1 MHz to 500 MHz

≤–65 dBmV

500 MHz to 1.0 GHz

≤–60 dBmV

1.0 GHz to 1.5 GHz

≤–53 dBmV

All specifications apply over 0 °C to +55 °C. The analyzer will meet its specifications five minutes after it is turned on,
when the analyzer is within one year of calibration cycle, after two hours of storage within the operating temperature range,
and Auto Align All is selected. ITALICS = supplemental information, characteristics, typical performance, or nominal values.

1 Frequency reference error = (aging rate x period of time since adjustment + settability + temperature stability).

2 N = Harmonic mixing mode. N = 1 for E4411B and E4403B.

3 For reference level 0 to –50 dBm: input attenuation, 10 dB; 50 MHz; RBW, 3 kHz, VBW, 3 kHz; log range 0 to 50 dB;

sweep time coupled, signal input, 0 to –50 dBm; span,

≤–60 kHz.

4 Mixer Power Level (dBm) = Input Power (dBm) – Input Attenuator. (dB).

5 For RBW ≤30 kHz, maximum input signal amplitude must be ≤ reference level + 10 dB.

6 Settings are: reference level –25 dBm for E4411B, –20 dBm for E4403B and E4408B; input attenuation 10 dB; center frequency

50 MHz; resolution bandwidth 3 kHz; video bandwidth 3 kHz; span 2 kHz; sweep time coupled; signal at reference level.