Specifi cations
(continued)
Amplitude specifications (continued)
Supplemental information
Level display range (continued)
Trace functions
Clear/write, maximum hold,
minimum hold, average
Level measurement error
1 MHz to 7 GHz
Excluding input VSWR mismatch
± 1.3 dB
• 20 to 30 °C, 30 to 70% RH, peak
detector, preamp off, input signal
-50 to 0 dBm, 95% percentile
• Swp Time Rule is set to Accuracy
• Adds additional ± 0.3 dB when Swp
Time Rule is set to Speed
7 to 13.6 GHz
± 1.6 dB
Reference level
2
Setting range
-100 to +30 dBm
Steps of 1 dB
Setting resolution
Log scale
0.01 dB
Linear scale
Same as log (2.236 µV to 7.07 V)
Accuracy
0
RF input VSWR (at tuned frequency)
10 MHz to 3 GHz
1 MHz to 7 GHz
< 1.5:1
Nominal, ≥ 10 dB attenuation
7 to 13.6 GHz
< 2:1
Spurious response
Second harmonic
distortion
< -65 dBc, typical < -70 dBc,
50 MHz to 7 GHz
Mixer signal level at -30 dBm, input
attenuation 0 dB, preamp off,
20 to 30 °C
< -80 dBc, typical < -90 dBc,
7 to 13.6 GHz
Third order intermodulation
distortion (third order
intercept)
+8 dBm, typical +9 dBm,
50 to 300 MHz
Two -20 dBm tones at input mixer,
spaced by 100 kHz, input attenuation
0 dB, preamp off, reference level
≥ -30 dBm, 20 to 30 °C
+9 dBm, typical +11 dBm,
300 MHz to 8 GHz
+10 dBm, typical +12 dBm,
8 to 13.6 GHz
Input related spurious
< -60 dBc, typical < -70 dBc
• -30 dBm signal at input mixer, span
< 2.9 GHz
• Exception:
-55 dBc (2 x F1 = center frequency
-5,890 MHz, 7 GHz < center frequency
< 10 GHz, with F1 input frequency)
Inherent residual response
< -95 dBm, typical -110 dBm,
1 MHz to 7 GHz
Input terminated and 0 dB RF
attenuation, preamplifier off
< -85 dBm, typical -93 dBm,
7 to 13.6 GHz
2. Reference level only affects the display not the measurement, so trace data markers do not cause additional errors in measurement results.