4
Specifi cations
(continued)
Amplitude specifications
Supplemental information
Measurement range
1 to 500 MHz
Displayed average noise level (DANL) to
+10 dBm
Preamp off
500 MHz to 13.6 GHz
Displayed average noise level (DANL) to
+20 dBm
Input attenuator range
0 to 50 dB, in 5 dB steps
Maximum safe input level
Average continuous
power
+30 dBm, 3 minutes maximum
Input attenuator setting ≥ 20 dB, 1 MHz
to 13.6 GHz
DC voltage
± 50 VDC maximum
Displayed average noise level
1
Preamp off
1 to 10 MHz
-125 dBm, typical -140 dBm
Reference level ≤ -50 dBm
10 MHz to 3 GHz
-137 dBm, typical -142 dBm
3 to 7 GHz
-135 dBm, typical -140 dBm
7 to 10 GHz
-139 dBm, typical -142 dBm
10 to 13.6 GHz
-137 dBm, typical -140 dBm
Preamp on
1 to 10 MHz
-140 dBm, typical -156 dBm
Reference level ≤ -70 dBm
10 MHz to 3 GHz
-150 dBm, typical -154 dBm
3 to 6 GHz
-145 dBm, typical -150 dBm
6 to 13.6 GHz
-151 dBm, typical -155 dBm
Level display range
Log scale
10 to 100 dB, 10 divisions displayed, 1, 2,
5, 10 dB/division
Linear scale
0 to 100%, 10 divisions displayed
Scale units
dBm, dBmV, dBμV, W, V, dBmV EMF,
dBμV EMF, V EMF
Sweep (trace) points
461
Marker level readout
resolution
Log scale
0.01 dB
Linear scale
≤ 1% of signal level
Nominal
Detectors
Normal, positive peak, sample, negative
peak, average (video, RMS, voltage)
Number of traces
4
1. RMS detector, trace averaging > 40, 0 dB input attenuation, input terminated 50 Ω, 1 kHz resolution bandwidth, normalized to 1 Hz, 20 to 30 °C.