beautypg.com

Maximum safe input level, 1 db gain compression, Displayed average noise level (danl) – Atec Agilent-E4401B-02B-04B-05B-07B User Manual

Page 3: Display range, Marker readout resolution, Frequency response, Input attenuation switching uncertainty at 50 mhz, Absolute amplitude accuracy

background image

3

Maximum safe input level

Average continuous power

(input attenuator

15 dB)

E4401B

+30 dBm (1 W)

E4401B (75

Option 1DP)

+75 dBmV (0.4 W)
(input attenuator

5 dB)

E4402B/04B/05B/07B

+30 dBm (1 W)

Peak pulse power

(input attenuator

30 dB)

E4401B

+30 dBm (1 W)

E4401B (75

Option 1DP)

+75 dBmV (0.4 W)

E4402B/04B/05B/07B

+50 dBm (100 W)

dc

E4401B, E4402B

100 Vdc

E4401B (75

Opt. 1DP)

100 Vdc

E4402B (Option UKB)

0 Vdc (dc coupled)
50 V (ac coupled)

E4404B, E4405B

0 Vdc (dc coupled)
50 V (ac coupled)

E4407B

0 Vdc

1 dB gain compression

(total power at input mixer

5

)

50 MHz to 6.7 GHz

0 dBm

6.7 GHz to 13.2 GHz

–3 dBm

13.2 GHz to 26.5 GHz

–5 dBm

Displayed Average Noise Level (DANL)

(dBm)

(Input terminated, 0 dB attenuation, sample detector)
1 kHz RBW; 30 Hz VBW
10 Hz RBW; 1 Hz VBW (Option 1DR)

1 kHz RBW 10 Hz RBW 10 Hz RBW

10 Hz RBW

(Option 1DR) (Option 1DR) (Option 1DR)

(w/preamp

(w/preamp

Option 1DS) Option 1DS)

Typical

E4401B
400 kHz to 1 MHz

–115

–134

–150

–155

1 MHz to 500 MHz

–119

–138

–154

–156

500 MHz to 1 GHz

–117

–136

–152

–156

1 GHz to 1.5 GHz

–114

–133

–150

–155

E4402B
30 Hz to 9 kHz

22

na

–93 na

na

(Option UKB)

9 kHz to 100 kHz

22

na

–109 na

na

100 kHz to 1 MHz

22

na

–135 na

na

1 MHz to 10 MHz

22

–117

–136

na

–152

10 MHz to 1 GHz

–117

–136

–152

19

–156

1 GHz to 2 GHz

–116

–135

–153

19

–156

2 GHz to 3 GHz

–114

–133

–151

19

–154

E4404/05B/07B
30 Hz to 9 kHz

22

na

–93 na

na

(Option UKB)

9 kHz to 100 kHz

22

na

–109 na

na

100 kHz to 1 mHz

22

na

–135 na

na

1 MHz to 10 MHz

22

–117

–137

na

–155

10 MHz to 1 GHz

–116

–135

–151

19

–157

1 GHz to 2 GHz

–116

–135

–151

19

–155

2 GHz to 3 GHz

–112

–131

–149

19

–152

3 GHz to 6 GHz

–112

–131

na

–138

6 GHz to 12 GHz

–111

–130

na

–137

12 GHz to 22 GHz

–107

–126

na

–134

22 GHz to 26.5 GHz

–106

–125

na

–132

E4407B (Option AYZ)
External mixer

6

–134 +

–153 +

na

na

external mixer external mixer

conversion conversion
loss loss

Display range

Log scale

0.1, 0.2, 0.5 dB/division and 1 to 20 dB/
division in 1dB steps;
ten divisions displayed.

RBW = 1 kHz

0 to -85 dB from reference level is
calibrated

RBW = 300 Hz (Option 1DR) 0 to -120

13

dB from reference level

is calibrated

Linear scale 10 divisions
Scale units

dBm, dBmV, dBµV, Volts, and Watts

(Option BAA)

add Hz

Marker readout resolution

Log scale

0 to –85 dB

0.04 dB

0 to –120 dB (Option 1DR)

0.04 dB

Linear scale

0.01% of reference level

Fast sweep times for zero span (Option AYX)
(sweeptimes

sweep points –1/100 kHz)

Log scale

0 to –85 dB

0.3 dB

Linear

0.3% of reference level

Frequency response

(10 dB input attenuation)

Absolute

7

/Typical

Relative flatness

8

E4401B

9 kHz to 1.5 GHz

±0.5 dB

na

±0.5 dB

E4402B/04B/05B/07B

30 Hz to 3 GHz

6

±0.5 dB

na

±0.5 dB

(Option UKB)

9 kHz to 3.0 GHz

±0.46 dB ±0.14 dB

±0.5 dB

3.0 GHz to 6.7 GHz

±1.5 dB

±0.38 dB

±1.3 dB

6.7 GHz to 13.2 GHz

±2.0 dB

±0.68 dB

±1.8 dB

13.2 GHz to 26.5 GHz ±2.0 dB

±0.86 dB

±1.8 dB

Input attenuation switching uncertainty at 50 MHz

Attenuation setting

0 dB to 5 dB

±0.3 dB

10 dB

reference

15 dB

±0.3 dB

20 to 60 dB (E4401B)

±(0.1 dB + 0.01 x attenuator setting)

20 to 65 dB

±(0.1 dB + 0.01 x attenuator setting)

Absolute amplitude accuracy

Typical

At reference settings

15

±0.34 dB

±0.13 dB

E4401B

±0.30 dB

±0.10 dB

Preamp on

16

(Option 1DS)

±0.37 dB

±0.14 dB

External mixer (Option AYZ)

IF INPUT absolute amplitude
accuracy + external mixer
conversion loss accuracy

17

Overall amplitude accuracy

9

±(0.54 dB + absolute frequency

response)