Atec Agilent-E4401B-02B-04B-05B-07B User Manual
Esa-e series spectrum analyzers, Frequency specifications, Data sheet
All specifications apply over 0 °C to + 55 °C unless otherwise noted
and are covered by the product warranty. The analyzer will meet its
specifications when: it’s within the one year calibration cycle, AUTO
ALIGN [ALL] is selected, stored a minimum 2 hours within the oper-
ating temperature range, turned on for at least 5 minutes, Align Now
RF has been run once every 24 hour period. Characteristics describe
product performance that is useful in the application of th product,
but is not covered by the product waranty. Typical performance is
beyond specifications that 80% of the units exhibit 95% confidence
level over 20 to 30°C not including measurement uncertainty and is
not covered by the product warranty.
Frequency specifications
Frequency range
E4401B
50
Ω
9 kHz to 1.5 GHz
75
Ω
1 MHz to 1.5 GHz
E4402B
9 kHz to 3.0 GHz
dc coupled (Option UKB)
30 Hz
6
to 3 GHz
ac coupled (Option UKB)
100 kHz to 3 GHz
E4404B
dc coupled
9 kHz to 6.7 GHz
dc coupled (Option UKB)
30 Hz
6
to 6.7 GHz
ac coupled
100 kHz to 6.7 GHz
Band
0
9 kHz to 3.0 GHz
(Option UKB)
100 Hz to 3.0 GHz
1
2.85 GHz to 6.7 GHz
E4405B
dc coupled
9 kHz to 13.2 GHz
dc coupled (Option UKB)
30 Hz
6
to 13.2 GHz
ac coupled
100 kHz to 13.2 GHz
Band
N
4
0
1–
9 kHz to 3.0 GHz
0
(Option UKB)
30 Hz
6
to 3.0 GHz
1
1–
2.85 GHz to 6.7 GHz
2
2–
6.2 GHz to 13.2 GHz
E4407B
Internal mixing
9 kHz to 26.5 GHz
dc coupled (option UKB)
30 Hz6 to 26.5 GHz
ac coupled (option UKB)
10 MHz to 26.5 GHz
Band N
4
0
1–
9 kHz to 3.0 GHz
0 (option
UKB)
30
Hz6 to 3.0 GHz
1
1–
2.85 GHz to 6.7 GHz
2
2–
6.2 GHz to 13.2 GHz
3
4–
12.8 GHz to 19.2 GHz
4
4–
18.7 GHz to 26.5 GHz
External mixing (Option AYZ)
18 GHz to 325 GHz
Agilent E4401B, E4402B, E4404B,
E4405B, and E4407B
Data Sheet
ESA-E Series Spectrum Analyzers
These specifications apply to the Agilent
Technologies E4401B, E4402B, E4404B,
E4405B, and E4407B spectrum analyzers.