beautypg.com

Amplitude specifications, Frequency reference, Frequency readout accuracy – Atec Agilent-E4401B-02B-04B-05B-07B User Manual

Page 2: Marker frequency counter, Frequency span, Sweep time, Sweep (trace) point range, Resolution bandwidth, Video bandwidth range, Stability

background image

2

Frequency reference

(Option 1D5)

Aging

±2

×

10

–6

/year

±1

×

10

–7

/year

Temperature stability

±5

Ч

10

–6

±1

Ч

10

–8

(20 to 30 °C)

Settability

±5

Ч

10

–7

±1

Ч

10

–8

Frequency readout accuracy

(Start, Stop, Center, Marker)

±(frequency indication x
frequency reference error

1

+ span

accuracy +15% of RBW + 10 Hz +
1 Hz x N

4

)

Marker frequency counter

2

Accuracy

3

±(marker frequency

×

frequency

reference error

1

+ counter

resolution)

Counter resolution

Selectable from 1 Hz to 100 kHz

Frequency span

Range

0 Hz (zero span), 100 Hz to the
maximum frequency range of
the analyzer

Resolution

2 Hz

×

N

4

Accuracy
(>2000 sweep points)

±0.5% of span

Sweep time

Range

Span >0 Hz

1 ms to 4000 s

Span = 0 Hz

10 µs to 4000 s

(Option AYX)

50 ns to 4000 s

(Option B7D)

25 ns to 4000 s

Accuracy

±1%

Sweep trigger

Free Run, Single, Line, Video,
External, delay, Offset,
Gate (Option 1D6), and TV
(Option B7B)

Delay trigger range

1 µs to 400 s

Sweep (trace) point range

101 to 8192

Span = 0 Hz

2 to 8192

Resolution bandwidth

1 kHz to 5 MHz (–3 dB) in 1-3-10
sequence.
9 kHz and 120 kHz (–6 dB) EMI
bandwidths.

Option 1DR

Adds 10, 30, 100, and 300 Hz (–3 dB)
bandwidths and 200 Hz (–6 dB)
EMI bandwidth.
(for spans

5 MHZ)

Accuracy

1 kHz to 3 MHz

±15%

5 MHz

±30%

10 Hz to 300 Hz (Option 1DR)

±10%

Selectivity (characteristic)

–60 dB/–3 dB
10 Hz to 300 Hz

<5:1

6

digital, approximately

Gaussian shape

1 kHz to 5 MHz

<15:1

6

synchronously tuned four

poles, approximately Gaussian
shape

Video bandwidth range

30 Hz to 3 MHz

6

in 1-3-10

sequence

Option 1DR

Adds 1 Hz, 3 Hz

,

and 10 Hz

(for RBW

1 kHz)

Stability

Noise sidebands (1 kHz RBW, 30 Hz VBW and sample detector)
Offset from CW signal

Typical

E4401B

1 kHz

na

–79 dBc/Hz (Option 1D5)

10 kHz

–93 dBc/Hz

–95 dBc/Hz

20 kHz

–100 dBc/Hz

–102 dBc/Hz

30 kHz

–104 dBc/Hz

–106 dBc/Hz

100 kHz

–113 dBc/Hz

–116 dBc/Hz

E4402/04/05/07B

1 kHz

na

-78 dBc/Hz (Option 1D5)

10 kHz

-90 dBc/Hz

21

-94 dBc/Hz

21

20 kHz

-100 dBc/Hz

21

-105 dBc/Hz

21

30 kHz

-106 dBc/Hz

21

-112 dBc/Hz

21

100 kHz

-119 dBc/Hz

21

-122 dBc/Hz

21

1 MHz

-125 dBc/Hz

21

-127 dBc/Hz

21

5 MHz

-127 dBc/Hz

21

-129 dBc/Hz

21

10 MHz

-131 dBc/Hz

21

-136 dBc/Hz

21

Option 120

1 MHz

-133 dBc/Hz

21

-136 dBc/Hz

21

5 MHz

-135 dBc/Hz

21

-139 dBc/Hz

21

10 MHz

-137 dBc/Hz

21

-141 dBc/Hz

21

Residual FM

1 kHz RBW, 1 kHz VBW

150

×

N

4

Hz pk-pk in 100 ms

Option 1D5

100

×

N

4

Hz pk-pk in 100 ms

Option 1DR

10

×

N

4

Hz

6

pk-pk in 20 ms

Option 1DR and 1D5

2

×

N

4

Hz pk-pk in 20 ms

System-related sidebands

30 kHz offset from CW signal

–65 dBc + 20 Log N

4

Amplitude specifications

Amplitude range

Measurement range

Displayed Average Noise Level
(DANL) to maximum safe input level

Input attenuator range

E4401B

0 to 60 dB, in 5 dB steps

E4402B/04B/05B

0 to 65 dB (75 dB

6

), in 5 dB steps

E4407B

0 to 65 dB, in 5 dB steps