Amplitude specifications, Frequency reference, Frequency readout accuracy – Atec Agilent-E4401B-02B-04B-05B-07B User Manual
Page 2: Marker frequency counter, Frequency span, Sweep time, Sweep (trace) point range, Resolution bandwidth, Video bandwidth range, Stability
2
Frequency reference
(Option 1D5)
Aging
±2
×
10
–6
/year
±1
×
10
–7
/year
Temperature stability
±5
Ч
10
–6
±1
Ч
10
–8
(20 to 30 °C)
Settability
±5
Ч
10
–7
±1
Ч
10
–8
Frequency readout accuracy
(Start, Stop, Center, Marker)
±(frequency indication x
frequency reference error
1
+ span
accuracy +15% of RBW + 10 Hz +
1 Hz x N
4
)
Marker frequency counter
2
Accuracy
3
±(marker frequency
×
frequency
reference error
1
+ counter
resolution)
Counter resolution
Selectable from 1 Hz to 100 kHz
Frequency span
Range
0 Hz (zero span), 100 Hz to the
maximum frequency range of
the analyzer
Resolution
2 Hz
×
N
4
Accuracy
(>2000 sweep points)
±0.5% of span
Sweep time
Range
Span >0 Hz
1 ms to 4000 s
Span = 0 Hz
10 µs to 4000 s
(Option AYX)
50 ns to 4000 s
(Option B7D)
25 ns to 4000 s
Accuracy
±1%
Sweep trigger
Free Run, Single, Line, Video,
External, delay, Offset,
Gate (Option 1D6), and TV
(Option B7B)
Delay trigger range
1 µs to 400 s
Sweep (trace) point range
101 to 8192
Span = 0 Hz
2 to 8192
Resolution bandwidth
1 kHz to 5 MHz (–3 dB) in 1-3-10
sequence.
9 kHz and 120 kHz (–6 dB) EMI
bandwidths.
Option 1DR
Adds 10, 30, 100, and 300 Hz (–3 dB)
bandwidths and 200 Hz (–6 dB)
EMI bandwidth.
(for spans
≤
5 MHZ)
Accuracy
1 kHz to 3 MHz
±15%
5 MHz
±30%
10 Hz to 300 Hz (Option 1DR)
±10%
Selectivity (characteristic)
–60 dB/–3 dB
10 Hz to 300 Hz
<5:1
6
digital, approximately
Gaussian shape
1 kHz to 5 MHz
<15:1
6
synchronously tuned four
poles, approximately Gaussian
shape
Video bandwidth range
30 Hz to 3 MHz
6
in 1-3-10
sequence
Option 1DR
Adds 1 Hz, 3 Hz
,
and 10 Hz
(for RBW
≤
1 kHz)
Stability
Noise sidebands (1 kHz RBW, 30 Hz VBW and sample detector)
Offset from CW signal
Typical
E4401B
≥
1 kHz
na
≤
–79 dBc/Hz (Option 1D5)
≥
10 kHz
≤
–93 dBc/Hz
≤
–95 dBc/Hz
≥
20 kHz
≤
–100 dBc/Hz
≤
–102 dBc/Hz
≥
30 kHz
≤
–104 dBc/Hz
≤
–106 dBc/Hz
≥
100 kHz
≤
–113 dBc/Hz
≤
–116 dBc/Hz
E4402/04/05/07B
≥
1 kHz
na
≤
-78 dBc/Hz (Option 1D5)
≥
10 kHz
≤
-90 dBc/Hz
21
≤
-94 dBc/Hz
21
≥
20 kHz
≤
-100 dBc/Hz
21
≤
-105 dBc/Hz
21
≥
30 kHz
≤
-106 dBc/Hz
21
≤
-112 dBc/Hz
21
≥
100 kHz
≤
-119 dBc/Hz
21
≤
-122 dBc/Hz
21
≥
1 MHz
≤
-125 dBc/Hz
21
≤
-127 dBc/Hz
21
≥
5 MHz
≤
-127 dBc/Hz
21
≤
-129 dBc/Hz
21
≥
10 MHz
≤
-131 dBc/Hz
21
≤
-136 dBc/Hz
21
Option 120
≥
1 MHz
≤
-133 dBc/Hz
21
≤
-136 dBc/Hz
21
≥
5 MHz
≤
-135 dBc/Hz
21
≤
-139 dBc/Hz
21
≥
10 MHz
≤
-137 dBc/Hz
21
≤
-141 dBc/Hz
21
Residual FM
1 kHz RBW, 1 kHz VBW
≤
150
×
N
4
Hz pk-pk in 100 ms
Option 1D5
≤
100
×
N
4
Hz pk-pk in 100 ms
Option 1DR
≤
10
×
N
4
Hz
6
pk-pk in 20 ms
Option 1DR and 1D5
≤
2
×
N
4
Hz pk-pk in 20 ms
System-related sidebands
≥
30 kHz offset from CW signal
≤
–65 dBc + 20 Log N
4
Amplitude specifications
Amplitude range
Measurement range
Displayed Average Noise Level
(DANL) to maximum safe input level
Input attenuator range
E4401B
0 to 60 dB, in 5 dB steps
E4402B/04B/05B
0 to 65 dB (75 dB
6
), in 5 dB steps
E4407B
0 to 65 dB, in 5 dB steps