beautypg.com

Csi 2140 machinery health analyzer – Atec Emerson-CSI-2140 User Manual

Page 9

background image

CSI 2140 Machinery Health Analyzer

June 2013

www.assetweb.com

9

Frequency analysis

A/D converter

24 bits of precision.

Automatic Integrator Correction

Allows precise measurement of low frequency vibrations down to 0.2 Hz.

Averaging modes

Normal, exponential, peak hold, order tracking, negative averaging, synchronous time

Cursors Spectrum

Single, Harmonic, Moving Harmonic, Sideband, and Time/Frequency for waveform.

Dynamic range

Converter has 120 dB dynamic range.

Frequency range

DC to 10 Hz minimum, DC to 80 kHz maximum

Frequency Units

Hz, CPM, Orders.

Full-scale range

Accelerometer input: 0-20V, Volts input -20V + 20V

Noise floor

Typically less than 20 μV for a 400-line spectrum at 1000 Hz maximum frequency.

Number of averages

5,000 in Route mode, 10,000 in Job mode.

Resolution

100, 200, 400, 800, 1600, 3200, 6400, or 12800 lines of resolution. True Zoom provides
effective resolution of up to 300,000 lines.

Response

Flat to DC for non-integrated and DC-coupled signals; optional AC coupling -3 dB at 1 Hz

Scaling

Linear or Log, both X and Y.

Windows

Hanning or uniform

Data storage capacity

Internal memory

1 GB

External memory

SD (Secure Digital). Virtually unlimited memory with off the shelf SDHC SD cards up to
32 GB.

Data analysis speed

400 line, 1000 HZ spectrum

67% overlap 6 avg/sec

1600 line, 1000 Hz spectrum

67% overlap 3 avg/sec