A series only – Atec Agilent-E8251A User Manual
Page 9
9
LF out
Output
Internal1 or internal2. Also provides monitoring of
internal1 or internal2 when used for AM, FM,
or FM.
Amplitude
0 to 3 Vpeak, nominal into 50
Ω
Output impedance
50 W, nominal
Swept sine mode:
(frequency, phase continuous)
Operating modes
Triggered or continuous sweeps
Frequency range
1 Hz to 1 MHz
Sweep rate
0.5 Hz to 100 k sweeps/s, equivalent to
sweep times 10 us to 2 s
Resolution
0.5 Hz (0.5 sweep/s)
Pulse modulation
≥ 500 MHz to ≤ 3.2 GHz
16
> 3.2 GHz
Power range
Internally leveled
0 to +10 dBm
0 to +10 dBm
With option 1E1
–110 to +10 dBm
–110 to +10 dBm
On/off ratio
80 dB typical
80 dB
Rise/fall times (T
r
, T
f
)
100 ns typical
10 ns (6 ns typical)
Pulse width
Internally leveled
≥ 2 µs typical
≥ 1µs
ALC Off
≥ 0.5 µs typical
≥ 20 ns typical
Repetition freq
Internally leveled
10 Hz to 250 kHz typical
10 Hz to 500 kHz typical
ALC Off
dc to 1 MHz typical
dc to 10 MHz typical
Level accuracy (relative to CW)
Internally leveled
±0.5 dB
±0.4 dB (±0.15 typical)
ALC Off with power search
17
±0.5 dB typical
≤ 20 GHz ±0.8 dB typical
≤ 40 GHz ±1.2 dB typical
Width compression
±50 ns typical
±5 ns typical
Video feedthrough
18
< 200 mV typical
< 2 mV typical
Pulse delay
(ext input to RF output)
300 ns nominal
70 ns nominal
Pulse overshoot (V
or
)
< 10% typical
Input level
+1 Vpeak = RF On
Input impedance
50
Ω, nominal
16
For improved performance
≤ 3.2 GHz, special Option
HE6 is available. Contact your local Agilent Online
representative.
17
Power search is a calibration routine that improves
level accuracy in ALC-off mode. Unpulsed RF power
will be present typically up to 5 ms when executing
power search.
18
With attenuator in 0 dB position. Video feed-through
decreases with attenuator setting.
A Series only