ISO compliant

This family of signal generators is manufactured in
an ISO-9001 registered facility in concurrence with
Agilent Technologies commitment to quality.

General

Power requirements

90 to 132 Vac 50 to 60 Hz, or 195 to 267
Vac 50 to 60 or 400 Hz, (automatically selected),
300 W maximum.

Operating temperature range

0 to 55°C

Storage temperature range

–40 to 71°C

Shock and vibration

Meets MIL-STD-28800E Type III, Class 3.

EMC

Conducted and radiated interference and immunity
meets IEC/EN 61326-1 and MIL-STD-461C Part 2,
RE02. Meets radiated emission requirements
of CISPR Pub 11/1997 Group 1 class A.

Storage registers

Memory is shared by instrument states, user data
files, sweep list files, and waveform sequences.
Depending on the number and size of these files,
up to 800 storage registers and 10 register
sequences are available.

Security

Display blanking.

Compatibility

Agilent 83550 Series millimeter heads

Self-test

Internal diagnostic routine tests most modules
(including microcircuits) in a preset condition. For
each module, if its node voltages are within
acceptable limits, then module “passes” the test.

Weight

< 22 kg (48 lb.) net, < 30 kg (68 lb.) shipping.

Dimensions

178 mm H x 426 mm W x 498 mm D
(7” H x 16.8” W x 19.6” D in.).

Front panel connectors

(All connectors are BNC female unless otherwise noted.)

RF output

Nominal output impedance 50

Ω.

For 20 GHz models

Precision APC-3.5 male, or Type-N with Option 1ED.

For 40 GHz models

Precision 2.4 mm male; plus 2.4-2.4 mm and
2.4-2.9 mm female adaptors also included.

ALC input

Used for negative external detector leveling.
Nominal input impedance 120 k

Ω,

damage level ±15 V.

LF output (PSG–A Series only)

Outputs the internally generated LF source.
Nominal output impedance 50

Ω.

External input 1 (PSG–A Series only)

Drives either AM, FM, or FM. Nominal input
impedance 50 or 600

Ω, damage levels are

5 Vrms and 10 Vpeak.

External input 2 (PSG–A Series only)

Drives either AM, FM, or FM. Nominal input
impedance 50 or 600

Ω, damage levels are

5 Vrms and 10 Vpeak.

Pulse/trigger gate input (PSG–A Series only)

Accepts input signal for external fast pulse
modulation. Also accepts external trigger pulse
input for internal pulse modulation. Nominal
impedance 50

Ω. Damage levels are 5 Vrms and

10 Vpeak.

Pulse video out (A series only)

Outputs a signal that follows the RF output in all
pulse modes. TTL-level compatible, nominal
source impedance 50

Ω.

L and A Series