Option 001 Phaselock/TCXO Timebase

Timebase Accuracy

Setability

< 0.01 ppm

Stability

± 1 ppm 0° - 50°

Aging

< 2ppm in first 30 days

(continuous operation)

0.1 pm/month

(after first 30 days)

External Reference Input
Lock Range

10 MHz ± 50 Hz

Level

-10 dBm to + 15 dBm
+25 dBm or 10 Vpp max
input

Impedance

50

± 2%, 42 Vpk isola-

tion to earth

Lock Time

< 2 seconds

Internal Reference Output
Frequency

10 MHz

Level

> 1 Vpp into 50

Phase Offset
Range

+ 360° to - 360°

Resolution

0.001°

Accuracy

25 ns

Trigger Output
Level

5V zero-going pulse

Pulse Width

> 2µs typical

Fanout

Capable of driving up to
three 33120As

Ordering Information

Agilent 33120A Function/Arb Generator
Opt. 001 Phase Lock/TCXO Timebase Option

General

Power Supply

110V/120V/220V/240V ±
10%

Power Line Frequency 45 Hz to 66 Hz and 360 Hz

to 440 Hz

Power Consumption

50VA peak (28 W aveage)

Operating Environment 0°C to 55°C

Storage Environment -40°C to 70°C

State Storage Memory Power Off state automati-

cally saved, 3 User
Configurable Stored
States

Interface

IEEE-488 and RS-232
standard

Language

SCPI - 1993, IEEE-488.2

Dimensions (W x H x D)

Bench top

254.4mm x 103.6mm x
374mm

Rack mount

212.6mm x 88.5mm x
348.3mm

Weight

4 kg (8.8 lbs)

Safety Designed to

UL-1244, CSA 1010,
EN61010

EMC Tested to

MIL-461C, EN55011,
EN50082-1

Vibration and Shock

MIL-T-28800, Type III,
Class 5

Acoustic Noise

30 dBa

Warm-up Time

1 hour

Warranty

3 years standard

254.4 mm

212.6 mm

374.0 mm

103.6

88.5 mm

348.3 mm