10
AGILENT 11722A SENSOR MODULE
SPECIFICATIONS
FREQUENCY RANGE: 100 kHz to 2.6 GHz.
POWER RANGE: +30 dBm (1 watt) to – 20 dBm (10
m
W).
INPUT SWR (connected to an 8902A):
<1.15, for RF Power measurements.
<1.33, for Tuned RF Level measurements, RF range 1 and 2.
<1.5, for Tuned RF Level measurements, RF range 3.
<1.33, for Tuned RF Level measurements, RF range 3 with
Special Function 1.9.
POWER SENSOR LINEARITY:
+2%, – 4%; +30 dBm to +20 dBm.
Negligible deviation, levels <+20 dBm.
CALIBRATION FACTORS:
Each 11722A sensor module is individually calibrated. The calibra-
tion factors are printed on the 11722A sensor module for easy ref-
erence.
CAL FACTOR UNCERTAINTY:
Frequency
RSS Uncertainty
Worst Case
(MHz)
Uncertainty
0.1
0.7 %
1.6%
0.3
0.7%
1.6%
1.0
0.8%
1.7%
3.0
0.8%
1.7%
10.0
0.9%
2.0%
30.0
0.9%
2.0%
50.0
0.0% (ref)
0.0% (ref)
100.0
1.1%
2.2%
300.0
1.1%
2.2%
1000.0
1.1%
2.2%
2600.0
1.2%
2.3%
Supplemental Characteristics:
MAXIMUM PEAK POWER: 100 Wpeak or 300 W ms per pulse.
INPUT IMPEDANCE: 50
Ω
nominal.
INPUT CONNECTOR: N-type male.
SWITCH LIFE: >1,000,000 switchings.
SWITCH ISOLATION: >90 dB.
WEIGHT: Net 0.8 kg (1.75 lb); Shipping 1.2 kg (2.6 lb).
DIMENSIONS: 51.2 mm H x 62.4 mm W x 1935 mm D
(2" x 2.5" x 76.2").
AGILENT 11792A SENSOR MODULE
SPECIFICATIONS
FREQUENCY RANGE:
RF Power measurements:
50 MHz to 26.5 GHz.
50 MHz to 18.0 GHz, Option 001.
POWER RANGE: +30 dBm (1 watt) to – 20 dBm (10
m
W).
INPUT SWR (connected to an Agilent 11793A):
<1.15, 1300 MHz
≤
f
c
.
<1.25, 1300 MHz c ≤ 18.0 GHz. <1.40, 18.0 GHz c ≤ 26.5 GHz. POWER SENSOR LINEARITY: CALIBRATION FACTORS: CAL FACTOR UNCERTAINTY: Frequency RSS Uncertainty Worst Case 2.0 GHz 2.3 4.6% 6.0 GHz 2.5 5.0% 10.0 GHz 2.9 5.7% 14.0 GHz 3.4 6.6% 18.0 GHz 3.7 6.9% 22.0 GHz 3.8 7.8% 26.5 GHz 4.1 8.3% Supplemental Characteristics: INPUT CONNECTOR: 3.5 mm male (N-type male, Option 001). INPUT IMPEDANCE: 50 Ω nominal. SWITCH LIFE: >1,000,000 switchings. WEIGHT: Net 0.8 kg (1.75 lb); Shipping 1.2 kg (2.6 lb). DIMENSIONS: 51.2 mm H x 62.4 mm W x 1935 mm D
+2%, – 4%; +30 dBm to +20 dBm.
Negligible deviation, levels <+20 dBm.
Each 11792A sensor module is individually calibrated. The calibra-
tion factors are printed on the 11792A sensor module for easy ref-
erence.
Uncertainty
(2" x 2.5" x 76.2").