SPECIFICATIONS
Model
62015B-15-90
62015B-30-50
62015B-60-25
62015B-80-18
62015B-150-10
Electrical Specifications
Output Ratings
Output Power
1350W
1500W
1500W
1440W
1500W
Output Voltage
1~15V
1~30V
1~60V
1~80V
1~150V
Output Current
1~90A
1~50A
1~25A
1~18A
1~10A
Line Regulation
0.1% FS
Load Regulation *1
1% FS
Programming Accuracy
1% FS
Measurement Accuracy
1% FS
Output Noise (20MHz)
Voltage Noise (P-P)
100mV
100mV
200mV
200mV
400mV
Voltage Ripple (rms)
30mV
30mV
50mV
50mV
100mV
Current Ripple (rms)
0.9A
0.5A
0.25A
0.18A
0.1A
Efficiency
> 87% @ full load
> 88% @ full load
Turn on over shoot voltage *2
5% of nominal output
Transient Response Time *3
< 5 ms
AC Input Voltage
Six Position Mainframe
187 ~ 250 Vac (3 Phase 4 Wire,
Δ
Connection ) or 323 ~ 437 Vac (3 Phase 5 Wire, Y Connection) /
45 ~ 65 Hz
Three Position Mainframe
187 to 250 Vac (single phase) / 45 ~ 65 Hz
Input Power Factor
> 0.98@ full load
Protection Function
OVP
Automatically shuts down at 115% of set value
Adjustment Range
1~16V
1~31V
1~65V
1~83V
1~155V
OCP
Current limit (0 ~ 100%) / OCP Shutdown at 115% of F.S.
OTP
Automatically shuts down if internal limit is reached
I/O Signal
Remote ON/OFF (I/P)
Dry contact (closed = enabled), vice versa
AUX Voltage
4 ~ 24V / 0.5A at mainframe (by trimmer adjust voltage)
DC OK Signal Type (O/P)
Dry contact (closed = enabled) (Error : OVP / OCP / OTP / AC Fault)
Programming Response Time *4 (Typical)
Rise Time (Full Load)
For a programmed 5% to 95% step in output voltage : 100ms
Rise Time (No Load)
For a programmed 5% to 95% step in output voltage : 100ms
Fall Time (Full Load)
For a programmed 95% to 5% step in output voltage : 40ms
Fall Time (No Load)
For a programmed 95% to 5% step in output voltage : 5s
Vout Setting
CAN Bus send command to DC module receiver : 1s
Measurement V & I
Under CAN command using fetch : 100ms
Delay Time
For output ON/OFF enable and disable (under CAN command) : 5s(Single Mainframe)
General Specifications
Remote Sensing
3V max. line loss compensation
Parallel Operation
Current Sharing (
±
5%)
Operating Temperature
0 ~ 50
℃
Humidity Range
0 ~ 90% RH. Non-condensing
Remote Interface
CAN Bus (optional)
Safety & EMC
CE
Dimension (H x W x D)
Mainframe : 175.6 x 443.9 x 466.2 mm / 6.91 x 17.48 x 18.35 inch (62000B-6-1)
Mainframe : 175.6 x 239.9 x 466.2 mm / 6.91 x 9.44 x 18.35 inch (62000B-3-1)
Module : 138.5 x 67.5 x 377.5 mm / 5.45 x 2.66 x 14.86 inch
Weight
Mainframe : 14 Kg / 30.8 lbs (62000B-6-1)
Mainframe : 8 Kg / 17.6 lbs (62000B-3-1)
Module : 4 Kg / 8.8 lbs
All specifications are subject to change without notice. Please visit our website for the most up to date specifications.
Note*1 :
For 50% step load variation with remote sense at maximum output voltage
Note*2 :
based on rise time of 100ms
Note*3 :
Time for the output voltage to recover within 1% of its rated for a load changed of 25%
Note*4 :
Six Position Mainframe through CAN