Display: 0.01 deg/div to 200 deg/div (0.00018 rad/div to 3.49
rad/div)

Accuracy (at 100 kHz, 25°C,
and Full Scale Input)

± 2.0°

Error

Input Level Relative

to Full Scale Input

Dynamic Accuracy

± .4 deg
± .2 deg
± .5 deg

± 1.5 deg
± 7.5 deg

0 dB to -10 dB

-10 dB to -50 dB
-50 dB to -60 dB
-60 dB to -80 dB

-80 dB to -100 dB

Reference Level Resolution

0.01°

Temperature Stability

Typically <± 0.05 deg/°C

Time Stability

Typically <± 0.05°/hour at 25°C

POLAR DISPLAY
CHARACTERISTICS

Range, Resolution, Display Units, Dynamic Accuracy, Frequency
Response, Uncertainty, Crosstalk, Reference Level, and Stability
specifications are the same as the corresponding magnitude and
phase characteristics.

Full Scale Magnitude Range

Absolute (A, B, R): 0.1 nV to 10 V
Ratio (A/R, B/R, A/B): 10

-10

to 10

20

REAL, IMAGINARY
DISPLAY
CHARACTERISTICS

Range, Dynamic Accuracy, Frequency Response, Uncertainty,
Crosstalk, Stability specifications are the same as the
corresponding magnitude and phase characteristics.

DELAY CHARACTERISTICS

Range

1 ps to 20,00s

Resolution

.01ns/div to 1000s/div

Normalized Accuracy

Dynamic Phase Accuracy +2nS
360 X Aperture [Hz]

Aperture Range

0.5% to 16% of frequency span

Reference Level

± 10

3

S

GENERAL DISPLAY CHARACTERISTICS

Traces

No. Traces

Two simultaneous traces may be present with a rectangular
graticule. One trace with polar or Smith graticules.
Markers: Each trace has one main marker and an offset marker.
Markers indicate data at corresponding trace coordinates in the
same units as used to set the Reference Level. Markers can be
used to modify certain display parameters. Marker resolution is
the same as horizontal display function.

Graticules

Rectangular Graticule

0% to 100% full scale deflection on 0.05% increments.