Atec Agilent-3577A User Manual
Atec Equipment
Hewlett Packard 3577A 5Hz - 200 MHz
Network Analyzer Specifications
SOURCE
Frequency
Range
5 Hz - 200 MHz
Resolution
0.001 Hz
Amplitude
± 5 X 10
-8
/day, 0 to 55°C
Range
+15 to -49 dBm (1.26 Vrms to 793m Vrms: 2dBV to -62 dBV)
into a 50Ω load
Resolution
0.1 dB
Accuracy
± 1 dB at +15 dBm and 100 kHz. Below +15 dBm, add the greater
of ± 0.02 dB/dB or 0.2 dB
Flatness
1.5 dBp-p from 5 Hz to 200 MHz
Impedance
50Ω; > 20 dB return loss at all levels
RF Output Connector
50Ω Type N female
Sweep Types
Linear, alternate, cw and log frequency; log amplitude
Sweep Time
100 ms/span to 200 ms/span for frequency sweep; 1 ms/step to 16
s/step for amplitude sweep
Sweep Modes
Continuous, single, manual
Trigger Modes
Free run, immediate, line, external
RECEIVER
Full Scale Input Level
-13 dBV from 10 kHz to 200 MHz with internal 20 dB attenuators
ON (0dBm at 50Ω)
INPUT CHARACTERISTICS
Frequency Range
5 Hz - 200 MHz
Inputs
3 (A, B, R)
Input Impedance
Selectable 50Ω with > 25 dB return loss, or 1 MΩ in parallel with
approximately 30 pF
Input Connectors
50Ω Type N female
Resolution Bandwidth
Selectable 1 kHz, 100 Hz, 10 Hz, or 1 Hz
30 kHz - 200 MHz (50Ω)
30 kHz - 20 MHz (1 MΩ)
Sensitivity (Due to noise and
internal crosstalk between
source and receiver inputs)
Resolution
Bandwidth
Internal 20 dB
Attenuator ON
Internal 20 dB
Attenuator OFF