Atec Agilent-3577A User Manual
Page 2
1 Hz
10 Hz
100 Hz
1 kHz
-110 dBm
-110 dBm
-105 dBm
-95 dBm
-130 dBm
-130 dBm
-125 dBm
-115 dBm
Crosstalk
> 100 dB isolation between inputs
Electrical Length/Reference
Plane Extension
Provides equivalent electrical line length, or delay at inputs A, B,
and R.
Range: -3 X 10
8
m to +3 X 10
8
m or +1 to -1 s
Resolution
5 digits or 0.1 cm (3.3 Ps) whichever is greater
Accuracy
± 0.1 cm or ± 0.02% whichever is greater
MAGNITUDE CHARACTERISTICS
Range
Full Scale Input to Sensitivity
Resolution
Marker: 0.002 dB (log); 5 digits (linear)
Display: 0.01 dB/div to 20 dB/div (log absolute); 0.01 dB/div to
200 dB/div (log ratio); 0.1 NV/div to 10 V/div (linear absolute);
10
-10
/div to 10
20
/div (linear ratio)
Display Units
dB, DBMS, div, V, and linear ratio
Accuracy
(at 100 kHz, 25°C, and Full Scale Input)
Absolute (A, B, R)
± 0.2 dB
Ratio (A/R, B/R, A/B)
± 0.15 dB (50Ω); ± 0.2 dB (1 MΩ)
Error
Resolution Bandwidth
1 kHz, 100 Hz, 10 Hz
1 Hz
Input Level Relative
to Full Scale Input
Dynamic Accuracy
± .04 dB
± .02 dB
± .05 dB
± .15 dB
± .75 dB
± .75 dB
± .04 dB
± .02 dB
± .05 dB
± .25 dB
± .75 dB
± 3.00 dB
0 dB to -10 dB
-10 dB to -50 dB
-50 dB to -60 dB
-60 dB to -80 dB
-80 dB to -90 dB
-90 dB to -100 dB
Frequency Response
(when driven from a 50Ω source and with a 50Ω receiver input impedance)
Absolute (A, B, R)
0.3 dBpp from 20 Hz to 20 MHz; 0.6 dBpp from 5 Hz to 200 MHz
Ratio (A/R, B/R, A/B)
0.3 dBpp from 20 Hz to 20 MHz; 0.4 dB from 5 Hz to 200 MHz
Reference Level
Range: -207 DBMS to +33 DBMS (-220 div to +20 div) (Log
absolute); -400 dB to +400 dB (log ratio); 0 V to 10 V (linear
absolute); 0 to 10
20
(linear ratio)
Stability
Temperature: Typically <± 0.02 dB/°C
Time: Typically <± 0.05 dB/hour at 25°C
PHASE CHARACTERISTICS
Range
± 180 degrees
Resolution
Marker: 0.005 deg (0.001 rad)