beautypg.com

Model – Atec Espec-SH Series User Manual

Page 11

background image

Operation mode

Program operation, Constant operation

7-segment LED display

Mechanical key input

9 steps/ 1 pattern (1 to 99 repetitions)

Thermocouple type T (Copper/ Copper-Nickel)

PID control

RS-485

Lithium battery, 1

Temp :−25 to +155℃ (SH-221, SU-221)

:−45 to +155℃ (SH-241・641,SU-241・641)

:−65 to +155℃ (SH-261・661,SU-261・661)

Humid : 0 to 100 %rh (SH only)

Tim

: 0 to 99 hours 59 minutes,

100 to 999 hours

Temp : 0.1℃

Humid : 1%rh (SH only)

Time : 1 minute

(in one hour unit for over 100 hours)

Input burn-out detection function

Upper and lower temperature (& humidity) limit alarm function

Self-diagnostic function (watchdog timer)

Alarm indication function

Power failure protection function

Timer function (automatic start/stop)

Refrigerator capacity automatic control function

Temp : 0.5℃ (Typ.)

Humid : ±2%rh (Typ.) (SH only)

Time : within 30 sec. per month

Display
Setting
Program capacity

Setting and

indication

ranges

Setting and

indication

resolution

Indication

accuracy *

Input
Control
Communication function

Battery

Auxiliary

functions

Model

ES-102

* At +23℃ ±5℃ ambient temperature

10

TEMPERATURE & HUMIDITY CONTROL RANGE (SH type only)

TEMPERATURE (& HUMIDITY) PROGRAM INDICATOR CONTROLLER

S□  

- □ □1

Temperature range 2: −20 to +150℃
     

4: −40 to +150℃

     

6: −60 to +150℃

Capacity 2: 22.5L

6: 64.0L

H: Low temperature & humidity chamber

U: Low temperature chamber

100

90

80

70

60

50

40

30

20

10

0

10

20

30

40

50

60

70

80

90 100

85℃

15℃

Temperature ℃

95%rh

Relative

Humidity

%rh

At +23℃ anbuebt temperature.

MODEL