beautypg.com

Burst module, pulse 3a/3b, Coupling matrix type cna – Atec EM-Test-UCS200M User Manual

Page 2

background image

EM TEST AG

Tel:

+41 (0)61 717 91 91

Sternenhofstr. 15

Fax:

+41 (0)61 717 91 99

ucs200m_eV115.doc

CH-4153 Reinach

email:

[email protected]

09.08.07

Switzerland

URL

http://www.emtest.com

Page 2/4

Burst module, Pulse 3a/3b

Electrical Fast Transient Simulator based on ISO 7637-2:2004

S

Coupling Matrix type CNA

Test Level Output

acc. to ISO 7637-2:2004
Test voltage

25V – 1’000V

± 10%

Rise time tr

5ns

± 1.5ns

Pulse duration td

100ns (+100/-0)ns

Verification

As per Annex D of ISO 7637-2:2004 into a
50

Ω and a 1,000Ω load

Source impedance

Zq = 50

Polarity

Pulse 3b positive and pulse 3a negative

Trigger Circuit

Trigger of bursts

Automatic, manual, external

Burst duration

T4 = 0.1ms - 999.9ms

Burst repetition rate

T5 = 10ms - 9,999ms

Spike frequency

f = 0.1kHz - 200kHz

Test duration

T = 0:01min - 999:59min or endless

Outputs

Direct

Via 50

Ω-coaxial connector

Coupling mode

To the + battery line

CRO trigger

5V trigger signal for oscilloscope

Test Routines

Quick Start

On-line adjustable parameters, easy to use

Standard Test Routines

acc. to ISO 7637-2:2004, level 1 - 4

User Test Routines

User Test Routines
Random burst release
Change voltage after T by

∆T

Change frequency after T by

∆f

CRO trigger

5V trigger signal for oscilloscope

Options

ACC

Capacitive coupling clamp acc. to ISO 7637 part 3

KW50

100:1 divider, 50

KW1000

400:1 divider, 1000

CA EFT kit

EFT/Burst verification kit

A6dB

6dB attenuator, 50

ITP

Immunity test probes (electrical field generation)

ITP/H

Immunity test probes (magnetic field generation

Technical data for standard CNA 50

DUT supply voltage

Max. 60V

DUT supply current

50 A

Inrush current capability

100A for 500ms

Dimension UCS & CN

19“ / 3HU

Technical data for optional CNA 100

DUT supply voltage

Max. 60V

DUT supply current

100 A

Inrush current capability

150A for 500ms

Dimension UCS & CN

19“ / 6HU

Technical data for optional CNA 150

DUT supply voltage

Max. 60V

DUT supply current

150 A

Inrush current capability

150 A

Dimension UCS & CN

19“ / 9HU

Technical data for optional CNA 200

DUT supply voltage

Max. 60V

DUT supply current

200 A

Inrush current capability

200 A

Dimension UCS & CN

19“ / 9HU

Input

DUT supply +/-

Simulator type VDS 200B & PFS 200B

Pulse 5, 7

Standard for one type LD 200x

Output

+/- DUT supply

Central DUT output

Coaxial output port

To connect the capacitive coupling clamp as
per ISO 7637 part 3

Interface

CN interface

To control the internal CN by external pulse
generators such as LD 200B

Option

CNA Ext

Extention to connect 3 additional generators