Burst module, pulse 3a/3b, Coupling matrix type cna – Atec EM-Test-UCS200M User Manual
Page 2
![background image](/manuals/318261/2/background.png)
EM TEST AG
Tel:
+41 (0)61 717 91 91
Sternenhofstr. 15
Fax:
+41 (0)61 717 91 99
ucs200m_eV115.doc
CH-4153 Reinach
email:
09.08.07
Switzerland
URL
http://www.emtest.com
Page 2/4
Burst module, Pulse 3a/3b
Electrical Fast Transient Simulator based on ISO 7637-2:2004
S
Coupling Matrix type CNA
Test Level Output
acc. to ISO 7637-2:2004
Test voltage
25V – 1’000V
± 10%
Rise time tr
5ns
± 1.5ns
Pulse duration td
100ns (+100/-0)ns
Verification
As per Annex D of ISO 7637-2:2004 into a
50
Ω and a 1,000Ω load
Source impedance
Zq = 50
Ω
Polarity
Pulse 3b positive and pulse 3a negative
Trigger Circuit
Trigger of bursts
Automatic, manual, external
Burst duration
T4 = 0.1ms - 999.9ms
Burst repetition rate
T5 = 10ms - 9,999ms
Spike frequency
f = 0.1kHz - 200kHz
Test duration
T = 0:01min - 999:59min or endless
Outputs
Direct
Via 50
Ω-coaxial connector
Coupling mode
To the + battery line
CRO trigger
5V trigger signal for oscilloscope
Test Routines
Quick Start
On-line adjustable parameters, easy to use
Standard Test Routines
acc. to ISO 7637-2:2004, level 1 - 4
User Test Routines
User Test Routines
Random burst release
Change voltage after T by
∆T
Change frequency after T by
∆f
CRO trigger
5V trigger signal for oscilloscope
Options
ACC
Capacitive coupling clamp acc. to ISO 7637 part 3
KW50
100:1 divider, 50
Ω
KW1000
400:1 divider, 1000
Ω
CA EFT kit
EFT/Burst verification kit
A6dB
6dB attenuator, 50
Ω
ITP
Immunity test probes (electrical field generation)
ITP/H
Immunity test probes (magnetic field generation
Technical data for standard CNA 50
DUT supply voltage
Max. 60V
DUT supply current
50 A
Inrush current capability
100A for 500ms
Dimension UCS & CN
19“ / 3HU
Technical data for optional CNA 100
DUT supply voltage
Max. 60V
DUT supply current
100 A
Inrush current capability
150A for 500ms
Dimension UCS & CN
19“ / 6HU
Technical data for optional CNA 150
DUT supply voltage
Max. 60V
DUT supply current
150 A
Inrush current capability
150 A
Dimension UCS & CN
19“ / 9HU
Technical data for optional CNA 200
DUT supply voltage
Max. 60V
DUT supply current
200 A
Inrush current capability
200 A
Dimension UCS & CN
19“ / 9HU
Input
DUT supply +/-
Simulator type VDS 200B & PFS 200B
Pulse 5, 7
Standard for one type LD 200x
Output
+/- DUT supply
Central DUT output
Coaxial output port
To connect the capacitive coupling clamp as
per ISO 7637 part 3
Interface
CN interface
To control the internal CN by external pulse
generators such as LD 200B
Option
CNA Ext
Extention to connect 3 additional generators