Micropulses – Atec EM-Test-UCS200N User Manual
Page 4
KARTA KATALOGOWA > UCS 200N-series > 20120209
PARAMETRY TECHNICZNE
MICROPULSES
MICROPULSE MODULE FOR TEST PULSES 1, 2A AND 6
As per ISO 7637-2:2011
Test voltage
U = 20V - 600V ± 10%
(peak voltage and polarity as per
selected standard)
Repetition rate
0.2s - 99.0s
ISO PULSE 1 (12V)
Rise time
1us +0%/-50% (10% - 90%)
Pulse duration
2ms ± 10% (10% - 10%)
Int. resistor
10ohm ± 10%
ISO PULSE 1 (24V)
Rise time
3us +0%/-50% (10% - 90%)
Pulse duration
1ms ± 10% (10% - 10%)
Int. resistor
50ohm ± 10%
ISO PULSE 2 (12V/24V)
Rise time
1us +0%/-50% (10% - 90%)
Pulse duration
50us ± 10% (10% - 10%)
Int. resistor
2ohm ± 10%
ISO PULSE 6 (12V) AS PER ISO 7637-1:1990
Rise time
60us +0%/-50% (10% - 90%)
Pulse duration
300us ± 10% (10% - 10%)
Int. resistor
30ohm ± 10%
AS PER ISO 7637-2 THE FOLLOWING STANDARDS CAN BE
COVERED
SAE J1113
GM 3097
BMW
Volkswagen
PSA
Chrysler PF 9326
DC 10614
Renault
FIAT
Mitsubishi
Honda
Ford ES-XW7T
MICROPULSES
TRIGGER
Automatic
Automatic release of the pulses
Manual
Manual release of a single pulse
External
External release of a single pulse
Battery switch
Off time, selectable; 0 - 10,000ms
OUTPUT
DUT supply +/-
Central DUT output
Coupling
To the battery +line
Decoupling
Via diode and battery supply switch
TEST ROUTINES
Quick Start
Immediate start; easy to use and fast
User Test routines
1. Custom made test routines
2. Change voltage after n pulses
Standard Test
routines
As per ISO 7637, SAE J1113, JASO
and manufacturer specifications
Service
Service, setup, self test
www.emtest.com
© EM TEST > STRONA 4/7