Other specifications – Atec Agilent-4339B_4349B User Manual

Page 6

background image

6

Other Specifications

Measurement parameters/ranges

Parameter

Range

Agilent 4339B

R (dc resistance)

10

3

Ω to 1.6 x

10

16

I (dc current)

60 fA to 100

µA

ps (surface

Refer to operation

resistivity)

manual

pv (volume

Refer to operation

resistivity)

manual

Agilent 4349B

R (dc resistance)

10

3

Ω to 10

15

I (dc current)

1 pA to 100

µA

Measurement conditions and functions
DC test voltage (4339B): 0 V to
1000 V, 0.1 V steps @ V

≤ 200V, 1.0 V

steps @ V > 200 V
DC test voltage (Agilent 4349B):
None supplied, use external power
supplies and voltage data entry
for resistance measurements.
Maximum of 5000 V input and 5
digit numerical entry.
Max current (Agilent 4339B):
10 mA @

≤ 100 V, 5 mA @ ≤ 250 V, 2

mA @

≤ 500 V, 1 mA @ ≤ 1 kV

Number of test channels:
4339B: 1 channel,
4349B: Option 4349B-700:4 ch
Option 4349B-001:2 ch
Ranging: Auto and hold
Trigger: Internal, manual, and
external
Delay time (trigger): 0 ms to
9999 ms in 1 ms steps
Test cable lengths: 2 meters
maximum
Measurement time (typical):
4339B: 10 ms / 30 ms / 390 ms
4349B: 9.5 ms / 28 ms / 98 ms /

397 ms

Other instrument functions
Error correction: Open (removes
errors due to parasitics).
Comparator: High, in, and low for
each of the test parameters.
Save/Recall: 10 instrument states
from non-volatile memory.
Contact check: Detects contact failure
for capacitive devices (2 ms).
GPIB: Agilent’s implementation of
IEEE 488 for control and data.
Handler interface: Negative logic
and isolated. Signals are
high/in/low, no contact, EOM, index,
alarm, keylock, ext. trigger.

Physical characteristics
Power: 90 – 132 Vac or 198 – 264 Vac.
47 Hz – 66 Hz. 45 VA (typical)
Operating
temperature/humidity
:
0 – 45 °C/

≤ 95% RH @ 40 °C.

Dimensions: 320(W) x 100(H) x
450(D) mm.
Weight: 6.5 kg (typical).